Service Information
Adjustment Procedure
20
Adjust Aberrations
Preparation
If the probe does not pass the performance verification procedure or you want to
optimize the performance of the probe, perform the following adjustment
procedure:
1. Insert a small screwdriver between the cover and the termination near the
2. Set up the oscilloscope as described below:
NOTE. When using a Wavetek 9100 calibrator with the scopecal option, no other
calibrators are required. The following setups are generic with nominal settings
indicated for the major adjustment steps.
CAUTION. To avoid damaging the probe, do not disconnect the probe termination
when measuring high currents. Leaving the probe unterminated can cause a high
voltage to develop in the secondary winding that may damage the current probe
transformer.
1. Connect the current probe calibration fixture to the fast rise output of the
2. Set the fast rise output to maximum (1 Vpp).
3. Connect the P6021 probe BNC connector to the termination.
4. Connect the P6021 termination to the oscilloscope Ch 1 input.
5. Set the P6021 termination sensitivity to 2 mA/mV.
part number, and gently pry up the top cover of the P6021 termination by
twisting the screwdriver. Leave the bottom cover on, since it must be in
place when the termination is in use.
Vertical mode
Horizontal mode
Trigger source
Trigger coupling
Trigger mode
Trigger slope
Trigger level
Ch 1 coupling
Ch 1 resistance
Volts/division
Time/division
Averages
calibration generator.
CH 1
Main
CH 1
AC
Peak-to-peak, auto
Positive
As required
DC
1 MΩ
2 mV
10 ns
5 to 10
P6021 Instruction Manual