BGA-3528 OKI METCAL, BGA-3528 Datasheet - Page 41

no-image

BGA-3528

Manufacturer Part Number
BGA-3528
Description
ASSEMBLY MICRO-OVEN 3500 SERIES
Manufacturer
OKI METCAL
Series
3500r
Datasheet

Specifications of BGA-3528

Accessory Type
Oven
For Use With/related Products
BGA and CSP Rework systems
Other names
BGA3528
10.
11.
marking. If the images are broken (not continuous) use the split vision adjustment knob to correct
and make the image continuous.
The circles should appear as one circle and all markings should be even from one side to the other.
If the gun sights are offset the unit will require calibration. The figure below on the left shows a
unit calibrated correctly and the figure on the right shows a unit out of calibration. If your
screen looks like the one on the right you must perform the factory calibration. Refer to Section
B, step 13 to perform beam splitter adjustment.
NOTE: Do not go to Step 11 if your unit is out of calibration. Perform adjustment before
Using split vision knob; rotate knob clockwise. Split the image bringing the outer lines in toward
the center of the screen. Zoom in on the image and center on the screen if necessary.
If the image on the monitor looks like the image above and on the left the magnification is in need
of adjustment. Refer to section D of the Factory Calibration. If the image on the monitor looks
like the image on the right follow sections C & D in the Factory Calibration.
proceeding to step 11.
ADDENDUM B
h
page
3