ISE40-T1-22 SMC Corporation of America, ISE40-T1-22 Datasheet - Page 2

no-image

ISE40-T1-22

Manufacturer Part Number
ISE40-T1-22
Description
SENSOR PRESSURE 1MPA NPN/ ANLG
Manufacturer
SMC Corporation of America
Series
ISE40r
Datasheet

Specifications of ISE40-T1-22

Pressure Type
Vacuum
Operating Pressure
1.45 PSI, 1 MPa
Port Size
Female, 1/8" (3.175mm) NPT
Output
Digital
Voltage - Supply
12 V ~ 24 V
Termination Style
Cable with Connector
Operating Temperature
0°C ~ 50°C
Package / Case
Module
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Other names
606-1005
Example of Internal Circuit and Wiring
Note 1) Equipped with unit switching function
Note 2) For ZSE40 (F)/ISE40-„-
Note 3) For ZSE40 (F)/ISE40-„-
Note 4) For ZSE40F (compound pressure) with “psi” indication, this is 0.03 to 0.04
Note 5) For ZSE40F (compound pressure) with “psi” indication, zero clear is in the
Specifications
ZSE40(F)
ISE40-„-22(L)-(M)
With analog output
ZSE40(F)
ISE40-„-30(L)-(M)
With auto shift input
16-2-18
Series
Rated pressure range
Operating pressure range/Set pressure range
Withstand pressure
Set pressure
resolution
Applicable fluid
Power supply voltage
Current consumption
Switch output
Repeatability
Hysteresis
Response time (With anti-chattering function)
Output short circuit protection
Display
Display accuracy
Indicator light
Analog output
Auto shift input
Environmental
resistance
Temperature characteristics
Port size
Lead wire
Weight
(Types without the unit switching function use SI units (kPa or MPa) only.)
psi.
range of ±0.01 psi.
Note 1)
Hysteresis mode
Window comparator mode
Note 2)
Enclosure
Ambient temperature range
Ambient humidity range
Withstand voltage
Insulation resistance
Vibration resistance
Impact resistance
Note 3)
ZSE40/ISE40
k g f/cm
mmH g
MPa
InH g
kPa
bar
psi
2
22
62
30
70
1 kΩ
6.8 kΩ
1.2 kΩ
±5% F.S. or less (in rated pressure range)
10 to 500 Hz at the smaller of amplitude 1.5 mm or acceleration 98 m/s
01/T1 types approx. 60 g, W1 type approx. 80 g, C4/C6/M5 types approx. 92 g (Each including 0.6 m lead wires)
Output impedance: Approx. 1 kΩ
ZSE40F (Compound pressure)
OUT2
(White)
DC (–)
(Blue)
DC (+)
(Brown)
Analog output
(Gray)
OUT1
(Black)
OUT2
(White)
DC (–)
(Blue)
DC (+)
(Brown)
Auto shift input
(Gray)
OUT1
(Black)
Linearity: ±1% F.S. or less
Output voltage: 1 to 5 V
–100.0 to 100.0 kPa
–100.0 to 100.0 kPa
0.02
2.5 ms or less (With anti-chattering function: 24 ms, 192 ms and 768 ms selections)
In a temperature range of 0 to 50°C, ±2% F.S. or less of pressure measured at 25°C
C4: With ø4 One-touch fitting, C6: With ø6 One-touch fitting, M5: M5 female threads
NPN or PNP 2 outputs Max. load current
Green LED (OUT1: Lights when ON), Red LED (OUT2: Lights when ON)
Operating: 0 to 50˚C, Stored: –10 to 60˚C (No condensation or freezing)
980 m/s
12 to 24 VDC
12 to 24 VDC
±2% F.S. ±1 digit or less (at ambient temperature of 25 ±3°C)
50 MΩ or more (at 500 VDC) between lead wires and body
No-voltage input (Reed or solid state), input 5 ms or more
01: R 1/8, M5 x 0.8, T1: NPT1/8, M5 x 0.8, W1: Rc 1/8
3 1/2 digit LED display (Sampling cycle: 5 times/sec.)
Operating/Stored: 35 to 85% RH (No condensation)
2
1000 VAC for 1 min. between lead wires and body
(100 G) in X, Y, Z directions 3 times each (De-energized)
500 kPa
5-wire oil resistant heavy-duty cord (0.15 mm
12 to 24 VDC ±10%, Ripple (p-p) 10% or less
0.001
0.001
0.1
0.1
1
Air, Non-corrosive/Non-flammable gas
ZSE40(F)
ISE40-„-62(L)-(M)
With analog output
ZSE40(F)
ISE40-„-70(L)-(M)
With auto shift input
Output voltage: 1 to 5 V ±2.5% F.S. or less (in rated pressure range)
Max. applied voltage: 30 VDC (With NPN output)
Residual voltage
ZSE40 (Vacuum pressure)
±0.2% F.S. ±1digit or less
Note)
When equipped with auto shift function, the following ranges can be set.
Fixed (3 digits)
10.0 to –101.3 kPa
0.0 to –101.3 kPa
Set pressure range
–100.0 to 100.0 kPa
55 mA or less
10.0 to –101.3 kPa
– 0.1 to 1.000 MPa
Variable
IP65
Yes
0.01
Output impedance: Approx. 1 kΩ
Note4)
2
Linearity: ±1% F.S. or less
: 80 mA
: 1 V or less (With 80 mA load current)
(10 G) in X, Y, Z directions for 2 hrs. each (De-energized)
1 k
6.8 k
1.2 k
–100.0 to 100.0 kPa
–101.3 to 101.3 kPa
–1.000 to 1.000 MPa
2
)
Setting range
ISE40 (Positive pressure)
–0.100 to 1.000 MPa
0.000 to 1.000 MPa
DC (+)
OUT1
(Black)
OUT2
(White)
Analog output
(Gray)
DC (–)
(Blue)
(Brown)
DC (+)
(Brown)
OUT1
(Black)
OUT2
(White)
DC (–)
(Blue)
Auto shift input
(Gray)
1.5 MPa
0.001
0.01
0.01
0.1
12 to 24 VDC
12 to 24 VDC

Related parts for ISE40-T1-22