FLUKE-66 Fluke Electronics, FLUKE-66 Datasheet - Page 17

THERMOMETER IR ADJ -25 TO 1100F

FLUKE-66

Manufacturer Part Number
FLUKE-66
Description
THERMOMETER IR ADJ -25 TO 1100F
Manufacturer
Fluke Electronics
Series
60r
Type
Handheld, Gunr
Datasheet

Specifications of FLUKE-66

Temperature Range
-25 ~ 1100°F (-32 ~ 600°C)
Temperature Display
C°/F°
Display Type
LCD
Input Type
Infrared; Probe
Features
Alarm, Backlight, Hold, Laser Sight, Memory, MIN/MAX/Dif/Avg
Probe Type
RTD
Battery Type
9 Volt
Includes
Case
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Other names
2148296
614-1040
63/66/68
Operating the Thermometer
Field of View
Make sure that the target is larger than the unit’s spot size. The
smaller the target, the closer you should be to it. See Figure 6.
ame004.eps
Figure 6. Field of View
Emissivity
Emissivity describes the energy-emitting characteristics of
materials. Most organic materials and painted or oxidized surfaces
have an emissivity of 0.95.
The 63 has a preset emissivity of 0.95. To compensate for
inaccurate readings that may result from measuring shiny metal
surfaces, cover the surface to be measured with masking tape or
flat black paint (< 148 C/300 F). Allow time for the tape or paint to
reach the same temperature as the surface beneath it. Measure
the temperature of the tape or painted surface.
13