FIELD CALIBRATION
The 343 has the ability to be field calibrated. Accuracy of +/-1°F
within the 30°F to 120°F range can be achieved by performing an ice
bath calibration.
The temperature of an ice bath is approximately 32°F (0°C). Fill a
plastic or metal container with crushed ice and add clean water to a
depth of at least 4 inches. Stir the ice and water for 2 to 3 minutes
prior to performing calibration to ensure the water is completely
chilled. Make certain there is plenty of ice in the mixture and
always use clean water. Distilled water works well.
1.
FOR FULL FACTORY CALIBRATION PLEASE RETURN YOUR THERMOMETER TO:
Step by Step Procedures
a.
b.
c.
d.
Connect both temperature probes to the 343.
Press and hold down the MIN/MAX and HOLD buttons and
turn the 343 on. CALF will be displayed and the 343 will
enter calibration mode and begin cycling between T1 and
T2.
Insert the probes into the ice bath and allow the readings to
stabilize.
Once the reading on both inputs has stabilized, press the
HOLD button. SAVE will be displayed and calibration is
complete.
TPI / ATTN: SERVICE DEPT.
9615 SW ALLEN BLVD SUITE 104
BEAVERTON, OR 97005
4