STM-16 ETC2 [List of Unclassifed Manufacturers], STM-16 Datasheet - Page 5

no-image

STM-16

Manufacturer Part Number
STM-16
Description
16 fold Spectroscopy Amplifier with Discriminators and Multiplicity Trigger
Manufacturer
ETC2 [List of Unclassifed Manufacturers]
Datasheet
mesytec
Memory List STM-16+:
ADR
0
1
3
4
5
7
8
9
11
12
13
15
The memory positions can be written with SE command and can be read with RE command.
A mirror page is available. It can be written with SM and read with RM and copied with CP.
The commands RST sets all thresholds to the value 32 and all gains to the value gain = 2.
The ON/OFF command makes the remote control active or inactive. The power up default is inactive. While
inactive the manual values from the front panel elements are set.
Identification code for STM (detected when running the scan bus command "SC") is IDC = 19
Example:
reset all channels of STM-16 module ID 7 on bus 0 to defined values:
RST 0 7
to set the channel 3 and 4 of module with ID 7 on bus 0 to gain = 1.5 and the threshold of channel 3 20% type:
SE 0 7 4 2
SE 0 7 5 128 // set threshold of channel 3 = address 5
activate remote control
ON 0 7
GAIN/THR
gain 1 and 2
threshold 1
threshold 2
gain 3 and 4
threshold 3
threshold 4
gain 5 and 6
threshold 5
threshold 6
gain 7 and 8
threshold 7
threshold 8
//set gain of channel 3 and 4 = address 4
fon: +49 - 89 / 456007-30, fax : +49 - 89 / 456007-50
wernher-von-braun- str. 1, 85640 putzbrunn, germany
info@mesytec.com - www.mesytec.com
ADR
16
17
19
20
21
23
24
25
27
28
29
31
mesytec GmbH & Co. KG
GAIN/THR
gain 9 and 10
threshold 9
threshold 10
gain 11 and 12
threshold 11
threshold 12
gain 13 and 14
threshold 13
threshold 14
gain 15 and 16
threshold 15
threshold 16
STM-16+