B25G210 Acopian Power Supplies, B25G210 Datasheet - Page 8

no-image

B25G210

Manufacturer Part Number
B25G210
Description
AC/DC POWER SUPPLY SNGL
Manufacturer
Acopian Power Supplies
Datasheet
High Voltage AC-DC
RACK MOUNTING
REGULATED
Output ranges:
Ideal for laboratory and instrumentation applications,
these rack mounting supplies have the same output
ratings and specifications as the modular supplies shown
on pages 4 and 5, but additionally feature calibrated
SPECIFICATIONS
Input Voltage: 105-125 VAC, 50-400 Hz, single phase.
Input Current:
Output Polarity: Positive output is standard. For negative
output, change first letter of model number from P to N.
Regulation (constant voltage operation):
Regulation (constant current operation):
Ripple: 0.05%, peak-to-peak.
Output Controls: Voltage and current may be controlled
by means of two 10-turn front panel adjustments with
locking vernier dials. Control linearity is 1% of full rated
output. Calibration accuracy is 1% of rated output
plus 1% of setting. (Remotely located 1000 ohm
potentiometers may alternately be used for output control.)
Metering: Voltmeter and ammeter are standard. Accuracy
is 2% of full scale.
Voltage Monitor Terminal: Permits remote monitoring of
output voltage, stepped down by ratio shown. Accuracy is
2% of maximum rated output voltage.
Current Monitor Terminal: Permits remote monitoring of
output current, at mV/mA ratio shown. Accuracy is 2% of
maximum rated output current.
Inhibit Terminal: Grounding inhibits output.
Input Protection: “Soft start” circuit minimizes start-up
power stresses.
• Shipped Within 9 Days
• Five Year Warranty
30 watt output ratings: 0.6A
60 watt output ratings: 1.2A
Line: ±0.05%
Load: ±0.05%
Line: ±0.1%
Load: ±0.1% plus 50 µA.
0 - 1,000 volts
It’s easy to find what you need at www.acopian.com or www.find a power supply.com
P.O. Box 638, Easton, PA 18044 • Phone: (610) 258-5441 • FAX: (610) 258-2842
to
0 - 30,000 volts
ten-turn controls (with locking vernier dials) for precisely
setting voltage and current. Voltmeter, ammeter and
handles are standard. An 8´ long shielded output cable is
included.
Output Programming: Output voltage and current may
be programmed from 0 to full rating by means of control
voltage inputs of 0 to +5.1 Vdc, ±2%.
Output Protection: Current regulation circuit protects
power supply from short circuits, overload, and arcing.
Response Time: Less than 5 mS for 100 µA load step
change.
Stability: 0.05% over eight hours, after 30 minute
warmup.
Temperature Coefficient: 200 PPM/°C = 0.02%/°C
(Typical).
Ambient Operating Temperature: –10 to +60°C.
No derating required.
Storage Temperature: –20 to +85°C.
Humidity: Maximum of 90% relative, non-condensing.
Connections:
OPTIONS
Terminal Strip Cover: Clips on AC input terminal strip.
To order, add suffix “M” to model number and $5.00 to
price.
230 Volt Input: All models can be alternately furnished
for operation on inputs of 210-250 VAC, 50-400 Hz,
single phase. To order, add suffix “–230” to model number
and add $40.00 to price. The “–230” option requires two
additional days.
Input, Control and Monitoring: Screw terminals.
Output: High voltage connector (Type varies with
model number). An 8´ shielded output cable, with
mating connector installed, is provided.
6