PM450 RSG Electronic Components GmbH, PM450 Datasheet

no-image

PM450

Manufacturer Part Number
PM450
Description
450-480 Watt Medical Switching Power Supplies
Manufacturer
RSG Electronic Components GmbH
Datasheet

Available stocks

Company
Part Number
Manufacturer
Quantity
Price
Part Number:
PM450CLA060
Manufacturer:
artesyn
Quantity:
1 000
Part Number:
PM450CLA060
Manufacturer:
MIT
Quantity:
20 000
Part Number:
PM450CLA120
Manufacturer:
VISHAY
Quantity:
12 000
Part Number:
PM450CLA120
Manufacturer:
MIT
Quantity:
20 000
Part Number:
PM450DV1A120
Manufacturer:
SAMSUNG
Quantity:
600 000
2-25
F E AT U R E S
‧ E N 6 1 0 0 0 - 3 - 2 c l a s s A a n d D c o m p l i a n t
‧ P o w e r F a c t o r 0 . 9 8 t yp i c a l
‧ O v e r v o l ta g e p r o t e c t i o n
‧ S h o r t - c i r c u i t p r o t e c t i o n
‧ T h e r m a l p r o t e c t i o n
‧ P o w e r F a i l D e t e c t ( P F D )
‧ 1 0 0 ﹪ b u r n - i n a t f u l l r a t e d l o a d
‧ R e m o t e s e n s e o n o u t p u t # 1 a n d o u t p u t # 2
‧ R e m o t e i n h i b i t – T T L h i g h d i s a b l e s o u t p u t
‧ C o m p l i a n t w i t h R o H S r e q u i r e m e n t s
DE S CR I PT IO N
The PM450 series comprising single and multiple output
models for 450-480 watts of continuous output power is
specially designed
support. They operate at 90 to 264VAC input voltage
without the need of a selector strap. The units are
constructed on a printed circuit board with a U-bracket for
mechanical support and heat sinking. A cover-and-fan
assembly can be added during manufacturing.
INPUT SPECIFICATIONS
Input voltage:
Input frequency:
Input current:
Earth leakage current:
O UT PUT S P EC IF IC AT IO N S
Output voltage/current: See rating chart
Total output power:
Ripple and Noise:
Overvoltage protection: Provided on output #1 only; set
Overcurrent protection: All outputs protected to short
Temperature coefficient: All outputs ± 0.04﹪ /
Transient response:
PFD signal:
Remote inhibit:
PROTEK
450-480 WATT MEDICAL SWITCHING POWER SUPPLIES
for medical applications not for life
See rating chart
2﹪ peak to peak maximum
at 115 – 140﹪ of its nominal
output voltage
circuit conditions
Maximum excursion of 4﹪ or
better on all models, recovering
to 1﹪ of final value within 500us
after a 25﹪ step load change
TTL logic high for normal
Operation and TTL logic low
upon loss of input power. This
signal appears at least 1ms prior
to master output dropping
5﹪ below its nominal value.
This signal also provides a
minimum delay of 100ms
after master output is within
regulation.
Requires an external TTL high
level signal to inhibit outputs for
standard models.
90 to 264VAC
47 to 63Hz
7.1A(rms)for 115VAC
3.5A(rms)for 230VAC
100uA max. @115VAC, 60Hz
165uA max. @ 230VAC, 50Hz
maximum
EN V IR O NM ENTAL S P EC IF IC AT IO N S
O p e r a t i n g t e m p e r a t u r e : 0 ℃ t o + 7 0 ℃
S t o r a g e t e m p e r a t u r e :
R e l a t i v e h u m i d i t y:
D e r a t i n g :
PM 450 SERIES
S af et y Stand a rd Ap p r ov al s:
GENERAL SPECIFICATIONS
Switching frequency: 60KHz ±10KHz
Power factor:
Efficiency:
Hold-up time:
Line regulation:
Inrush current:
Withstand voltage:
MTBF:
EMC Performance (EN60601-1-2)
EN55011:
EN61000-3-2:
EN61000-3-3:
EN61000-4-2:
EN61000-4-3:
EN61000-4-4:
EN61000-4-5:
EN61000-4-6:
EN61000-4-8:
EN61000-4-11:
TÜV EN60601-1
UL60601-1, CSA C22.2 No. 601.1
File No. E178020
0.98 typical
80﹪ minimum on all models
12 msec minimum at 110VAC
±0.2﹪ maximum at full load
40 amps @ 115VAC or 80 amps
@ 230VAC, at 25
4000VAC from input to output
1500VAC from input to ground
500VAC from output to ground
300,000 hours minimum at full load
at 25
MIL-HDBK- 217F
Class B conducted, Class A radiated
Harmonic distortion, Class A and D
Line flicker
ESD, ±8KV air and ±6KV contact
Radiated immunity, 3V/m
Fast transient/burst, ±2KV
Surge, ±1KV diff., ±2KV com.
Conducted immunity, 3Vrms
Magnetic field immunity, 3A/m
Voltage dips, 30% reduction for
500ms, 60% reduction for 100ms
and >95% reduction for 10ms
- 4 0 ℃ t o + 8 5 ℃
5 ﹪ t o 9 5 ﹪ n o n - c o n d e n s i n g
D e r a t e f r o m 1 0 0 ﹪ a t + 5 0 ℃
l i n e a r l y t o 5 0 ﹪ a t + 7 0 ℃
℃ mbient, calculated per
a
cold start
RoHS
( LV D )

Related parts for PM450

PM450 Summary of contents

Page 1

... WATT MEDICAL SWITCHING POWER SUPPLIES PROTEK The PM450 series comprising single and multiple output models for 450-480 watts of continuous output power is specially designed for medical applications not for life support. They operate 264VAC input voltage without the need of a selector strap. The units are constructed on a printed circuit board with a U-bracket for mechanical support and heat sinking ...

Page 2

... PM450-37 48V 0.5A NOTES: 1. Add suffix “B” for U-bracket format or “C” for enclosed format with fan control function, e.g. PM450-14C. 2. All outputs are floating. It can be connected externally for positive or negative output. 3. Output #1 can be adjusted within ±5% of their nominal voltage. 4. Output #3 can be adjusted within ±15% of their nominal voltage. ...

Page 3

... NEUTRAL GROUND +V1 - INHIBIT +S (V2) -S (V2) 0V (FAN) -Ve PM450 SERIES NOTES: 1. Dimensions shown in inch [mm]. 2. Tolerance 0.02 [0.5] maximum. 3. Input connector P1 is Dinkle DT-35-B01W-03 screws are M3, Nickel plated. 4. Connector P4 mates with Molex housing 50-37-5103 and pins 5263. 5. P2, P3: M*0.05 screw connections ...

Related keywords