ELECTRONIC COMPONENTS
DESCRIPTION
The PM201 series comprising single and multiple output
models for 150 to 200 watts of continuous output power is
specially designed for medical and ITE applications, not for
life-support. They operate at 90 to 264 VAC input voltage
without the need of a selector strap. All auxiliary outputs are
with magnetic amplifier linear regulator to keep regulation.
The units are constructed on a printed circuit board with a
U-bracket for mechanical support and heat sinking. A
cover-and-fan assembly can be added during manufacturing.
FEATURES
INPUT SPECIFICATIONS
Input voltage:
Input frequency:
Input current:
Earth leakage current:
OUTPUT SPECIFICATIONS
Output voltage/current:
Total output power:
Ripple and noise:
Overvoltage protection:
Overcurrent protection:
Temperature coefficient:
Transient response:
PFD signal:
ENVIRONMENTAL SPECIFICATIONS
Operating temperature:
Storage temperature:
Relative humidity:
Derating:
Cooling:
L o w s a f e t y g r o u n d l e a k a g e c u r r e n t
M e e t E N 5 5 0 11 a n d F C C C l a s s B
P o w e r F a c t o r 0 . 9 8 t y p i c a l
S h o r t - c i r c u i t p r o t e c t i o n
P o w e r F a i l D e t e c t ( P F D ) s i g n a l
1 0 0 % b u r n - i n a t f u l l r a t e d l o a d
O p t i o n a l c o v e r - a n d - f a n a s s e m b l y
R e m o t e i n h i b i t - T T L h i g h d i s a b l e s o u t p u t
C o m p l i a n t w i t h R o H S r e q u i r e m e n t s
RSG
■
RSG Electronic Components GmbH
Tel. +49 69 984047-0
Änderungen vorbehalten / subject to change without notice
90-264 VAC
47-63 Hz
3.20 A (rms) for 115 VAC
1.60 A (rms) for 230 VAC
90 uA max. @ 115 VAC, 60 Hz
150 uA max. @ 230 VAC, 50 Hz
See rating chart
200 watts maximum
2% peak to peak maximum
Provided on output #1 only;
set at 112-132% of its nominal output
voltage
All outputs protected to short circuit
conditions
All outputs ±0.04% /℃ maximum
Maximum excursion of 4%
or better on all models, recovering to 1%
of final value within 500 us after a
25% step load change
TTL logic high for normal operation and
TTL logic low upon loss of input power.
This signal appears at least 1ms prior to
+5 V output dropping 5% below its
nominal value. This signal also provides
a minimum delay of 100 ms after +5 V is
within regulation.
0℃ to +70℃
-40℃ to +85℃
5% to 95% non-condensing
Derate from 100% at +50℃
linearly to 50% at +70℃ 10.8 CFM
forced air provided on “C” version; 25
CFM forced air to be provided for “B”
version by user.
150-200 WATT MEDICAL SWITCHING POWER SUPPLIES
■
Fax +49 69 984047-77
■
Sprendlinger Landstr. 115
15
PM201 SERIES
■
SAFETY STANDARD APPROVALS
GENERAL SPECIFICATIONS
Switching frequency:
Efficiency:
Hold-up time:
Line regulation:
Inrush current:
Withstand voltage:
MTBF:
EMC Performance (EN60601-1-2: 2001)
EN55011/EN55022:
EN61000-3-2:
EN61000-3-3:
EN61000-4-2:
EN61000-4-3:
EN61000-4-4:
EN61000-4-5:
EN61000-4-6:
EN61000-4-8:
EN61000-4-11:
info@rsg-electronic.de
■
UL 60601-1, CSA C22.2 No. 601.1
File No. E178020
UL 60950-1
File No. E137410
TÜV EN60601-1
TÜV EN60950-1
D-63069 Offenbach/Germany
88-112 KHz
70% minimum on all models
20 msec minimum at 110 VAC
±0.5% maximum at full load
20 A @ 115 VAC or 40 A @ 230 VAC,
at 25℃ cold start
4000 VAC from input to output
1500 VAC from input to ground
500 VAC from output to ground
350,000 hours at full load at 25℃ ambient ,
calculated per MIL-HDBK-217F
Class B conducted, class B radiated
Harmonic distortion, class A and D
Line flicker
ESD, ±8 KV air and ±6 KV contact
Radiated immunity, 3 V/m
Fast transient/burst, ±2 KV
Surge, ±1 KV diff., ±2 KV com.
Conducted immunity, 3 Vrms
Magnetic field immunity, 3 A/m
Voltage dips, 30% reduction for 500 ms,
60% reduction for 100 ms and >95%
reduction for 10 ms
■
www.rsg-electronic.de
RoHS
( LV D )