FLUKE-190-204/AM Fluke Electronics, FLUKE-190-204/AM Datasheet - Page 66

SCOPEMETER 4CH 200MHZ COLOR AMER

FLUKE-190-204/AM

Manufacturer Part Number
FLUKE-190-204/AM
Description
SCOPEMETER 4CH 200MHZ COLOR AMER
Manufacturer
Fluke Electronics
Series
ScopeMeter® 190r
Type
Handheldr

Specifications of FLUKE-190-204/AM

Bandwidth
200MHz
Channels
4
Display Type
LCD - Color
Features
DMM, USB Interface, Save, Recall, Record, Trend
Probe Type
10:1 Voltage Probe (1 Blue, 1 Green, 1 Red, 1 Gray)
Sampling Rate (per Second)
1.25GS/s
Input Impedance
1M
Rise Time (typ)
1.7nS
Voltage - Input (max)
CAT III 1000V, CAT IV 600V
Voltage - Supply
7.2V Battery, 17.8VDC Adapter
Includes
Alligator Clips, Battery Charger/Adapter, Case, Hook Clips, Probe Set, Software, Test Leads, USB Cable
Lead Free Status / Rohs Status
Lead free / RoHS Compliant
Other names
614-1174
Fluke 190 series II
Users Manual
Setting Trigger Level and Slope
The Connect-and-View™ function enables hands-off
triggering to display complex unknown signals.
When your test tool is in manual range, do the following:
Automatic triggering assures a stable display of virtually
any signal.
From this point, you can take over the basic trigger
controls such as level, slope and delay. To optimize trigger
level and slope manually, do the following:
58
1
2
3
F2
F3
MANUAL
TRIGGER
AUTO
Perform an auto set.
at the top right of the screen.
Display the
Trigger on either positive slope or
negative slope of the chosen
waveform.
In Dual Slope Triggering ( X ) the
test tool triggers on both positive
slope and negative slope.
Enable the arrow keys for manual
trigger level adjustment.
TRIGGER
key labels.
AUTO
appears
Observe the trigger icon
position, trigger level, and slope.
At the bottom of the screen the trigger parameters are
displayed (See Figure 28). For example,
means that input A is used as the trigger source with a
positive slope.
When a valid trigger signal is found, the trigger key will be
lit and the trigger parameters appear in black.
When no trigger is found, the trigger parameters appear in
gray, ands the key light will be off.
4
Figure 28. Screen with all Trigger Information
Adjust the trigger level.
that indicates the trigger