Specifications of FLUKE-190-204/AM

Bandwidth
200MHz
Channels
4
Display Type
LCD - Color
Features
DMM, USB Interface, Save, Recall, Record, Trend
Probe Type
10:1 Voltage Probe (1 Blue, 1 Green, 1 Red, 1 Gray)
Sampling Rate (per Second)
1.25GS/s
Input Impedance
1M
Rise Time (typ)
1.7nS
Voltage - Input (max)
CAT III 1000V, CAT IV 600V
Voltage - Supply
7.2V Battery, 17.8VDC Adapter
Includes
Alligator Clips, Battery Charger/Adapter, Case, Hook Clips, Probe Set, Software, Test Leads, USB Cable
Lead Free Status / Rohs Status
Lead free / RoHS Compliant
Other names
614-1174
Working with Noisy Waveforms
To suppress high frequency noise on waveforms, you can
limit the working bandwidth to 20 kHz or 20 MHz. This
function smoothes the displayed waveform. For the same
reason, it improves triggering on the waveform.
To choose HF reject on for example input A, do the
following:
1
2
3
To suppress noise without loss of bandwidth,
use the average function or turn off Display
Glitches.
F4
A
ENTER
Display the
Open the
Jump to Bandwidth: and select
20kHz (HF reject) to accept the
bandwidth limitation.
Tip
INPUT A
INPUT A
menu.
key labels.
Using Mathematics Functions +, -, x, XY-mode
You can add (+), subtract (-), or multiply (x) two
waveforms. The test tool will display the mathematical
result waveform and the source waveforms.
The XY-mode provides a plot with one input on the
vertical axis and the second input on the horizontal axis.
The Mathematics functions perform a point-to-point
operation on the involved waveforms.
To use a Mathematics function, do the following:
1
2
3
F4
SCOPE
ENTER
Display the
Open the
Jump to Waveform: and Select
Mathematics... to open the
Mathematics menu.
Acquiring Waveforms
WAVEFORM OPTIONS
SCOPE
key labels.
menu.