FLUKE-190-202/AM/S Fluke, FLUKE-190-202/AM/S Datasheet - Page 6

ScopeMeter, 200 MHz, 2 Channels Plus DMM/Ext.input, With SCC-290 Kit

FLUKE-190-202/AM/S

Manufacturer Part Number
FLUKE-190-202/AM/S
Description
ScopeMeter, 200 MHz, 2 Channels Plus DMM/Ext.input, With SCC-290 Kit
Manufacturer
Fluke
Datasheet

Specifications of FLUKE-190-202/AM/S

Scope Type
Hand Held
Scope Channels
2
Bandwidth
200MHz
Meter Display Type
LCD Color
Sampling Rate
2.5GSPS
Input Impedance
1Mohm
Supply Voltage Min
600V
Supply Voltage Max
1kV
Rohs Compliant
NA
Display
153 mm Full Color LCD
Equipment Type
Oscilloscopes
Lead Free Status / Rohs Status
 Details
General Specifications
6 Fluke Corporation
Input voltage range
Rated maximum floating voltage
Maximum probe voltage
Maximum BNC input voltage
Maximum voltage on meter input
Memory save and recall
Memory locations (internal)
15 waveform memory locations
Two recording memories
External data storage
Screencopies
Volatility
Real-time clock
Case
Design
Drip and dust proof
Shock and vibration
Display size
Resolution
Contrast and brightness
Brightness
Mechanical data
Size
Weight (including battery)
Power
Line power
Battery power
Battery type (incl.) and capacity
[+opt. battery]
Battery charge indicator
Battery operating time
(with backlight low)
Battery charging time
Battery power saving functions
Safety
Compliance
ScopeMeter 190 Series II
15 waveform memories plus 2 recording memories
Stores Scope-trace waveform data (2 or 4 traces each) plus screen-copy plus corresponding setup
Each may contain:
• a 100 Screen Replay sequence, or
• a ScopeRecord Roll-mode recording (2 or 4 traces), or
• a TrendPlot recording of up to 4 measurements
• On PC, using FlukeView
• Direct storage on external flash memory drive (maximum 2 GB) through USB host port
• On PC, using FlukeView
• Internally (in instrument) which can be copied on to external flash memory drive as .BMP-file, through
Measurement data is initially stored in RAM, which is maintained by the main battery with a 30 seconds
back-up when battery is exchanged
When storing data, this is written in non-volatile flash-ROM
Provides date and time stamp information for ScopeRecord, for 100 Screen Replay sequences and for
TrendPlot recordings
Rugged, shock-proof with integrated protective holster. Handstrap and hangstrap included as standard
Kensington lock supported to lock down instrument when left unattended
IP 51 according to IEC529
Shock 30 g, vibration (sinusoidal) 3 g according to MIL-PRF-28800F Class 2
127 mm x 88 mm (153 mm/6.0 in diagonal) LCD
320 x 240 pixels
User adjustable, temperature compensated
200 cd/m
Re-chargeable double capacity Li-Ion battery (included). Battery swappable through easily accessible
Auto ‘power down’ with adjustable power down time; Auto ‘Display off’ with adjustable power down
Battery has built-in status indicator for use with external charger, next to battery status indicator on
USB host port
190-062
2½ hours using BP290; 5 hours using BP291
2
(safety designed banana input connectors)
Up to eight hours using BP291 (optional)
Up to four hours using BP290 (included),
typ. using power adapter, 90 cd/m
Mains adapter/battery charger BC190 included, version depending of country
[BP291 (4800 mAh) optional]
(maximum voltage between any contact and earth-ground voltage level)
(maximum voltage between standard 10:1 probe tip and reference lead)
CAT III 1000V/CAT IV 600V
BP290; 2400 mAh
2.1 kg (4.6 lb)
265 mm x 190 mm x 70 mm (10.4 in x 7.5 in x 2.8 in)
190-102
CAN/CSA C22.2, No. 61010-1-04, with approval;
Software, or
Software, or
battery door at the rear of the instrument
(maximum voltage on BNC input directly)
time; On-screen battery power indicator
EN61010-1-2001, Pollution Degree 2;
CAT III 1000V/CAT IV 600V
CAT III 1000V/CAT IV 600V
ANSI/ISA-82.02.01
instrument screen
CAT IV 300 V
2
190-202
UL61010B;
typical using battery power
Up to seven hours using BP291 (included)
190-104
BP291; 4800 mAh
Five hours BP291
2.2 kg (4.8 lb)
190-204