U1731B Agilent Technologies Test Equipment, U1731B Datasheet - Page 9

no-image

U1731B

Manufacturer Part Number
U1731B
Description
Orange Handheld LCR Meter
Manufacturer
Agilent Technologies Test Equipment
Type
LCR Meterr
Datasheet

Specifications of U1731B

Backlight
Yes
Capacitance Accuracy
Up to 0.7% + 5 lsd
Capacitance Range
0.1 pF to 199.99 mF
Color, Body
Orange
Display Type
Dual-Display LCD
Includes
Quick Start Guide, Alligator clip leads, 9 V Alkaline battery, Certifi cation of Calibration (CoC)
Interface
IR to USB
Resistance Range
0.001 Ω to 9.999 MΩ
Resolution
20,000 Counts
General Specifications
Parameter
Measurements
Tolerance mode
Test frequency (Accuracy =
±0.1% of actual test frequency)
Measuring circuit mode
Display
Backlight
Ranging mode
Test signal level
Measurement rate
Response time
Auto power-off
Power supply
Power consumption
Input protection fuse
Battery life
Low battery indicator
Operating environment
Storage environment
Temperature coefficient
Weight
Dimensions (H x W x D)
Safety and EMC compliance
Calibration
Warranty
U1731A/U1731B
L/C/R/D/Q
1%, 5%, 10%
Test frequency
setting
120 Hz
1 kHz
Inductance (L): Defaults to series mode
Capacitance/Resistance (C/R): Defaults to parallel mode
L/C/R : Maximum display 19999
D/Q: Maximum display 999 (Auto range)
Available for model U1732A/U1732B
Auto and Manual
~0.6 V
1 reading/s, nominal
~1 s/DUT (manual range)
~5 mins without operation
0.1 A/250 V
7 hours (typical) without backlight and based on new alkaline
0 °C to 40 °C; 0 to 70% relative humidity (R.H.)
–20 °C to 50 °C; 0 to 80% R.H. non-condensing
0.15 x (specified accuracy)/°C (0 °C to 18 °C or 28 °C to 40 °C)
330 g
184 mm x 87 mm x 41 mm
IEC 61010-1:2001/EN 61010-1:2001 (2
EN 61326-2-1:2006, ICES-001:2004, AS/NZS CISPR11:2004
One-year calibration cycle recommended
3 years for main unit
3 months for standard shipped accessories
9 V Alkaline battery (ANSI/NEDA 1604A or IEC 6LR61)
AC power adapter and cord available as options
~40 mA (on battery operation)
0.08 mA after auto power-off
RMS
will appear when the voltage drops below ~ 6.8 V
Actual test
frequency
120 Hz
1010 Hz
9
nd
Edition) Pollution Degree 2, IEC 61326-2-1:2005/
U1732A/U1732B
L/C/R/D/Q/q
1%, 5%, 10%, 20%
Test frequency
setting
100 Hz
120 Hz
1 kHz
10 kHz
Actual test
frequency
100 Hz
120 Hz
1010 Hz
9.6 kHz