U1731B Agilent Technologies Test Equipment, U1731B Datasheet - Page 7

no-image

U1731B

Manufacturer Part Number
U1731B
Description
Orange Handheld LCR Meter
Manufacturer
Agilent Technologies Test Equipment
Type
LCR Meterr
Datasheet

Specifications of U1731B

Backlight
Yes
Capacitance Accuracy
Up to 0.7% + 5 lsd
Capacitance Range
0.1 pF to 199.99 mF
Color, Body
Orange
Display Type
Dual-Display LCD
Includes
Quick Start Guide, Alligator clip leads, 9 V Alkaline battery, Certifi cation of Calibration (CoC)
Interface
IR to USB
Resistance Range
0.001 Ω to 9.999 MΩ
Resolution
20,000 Counts
[1] This reading can be extended up to 1999 MAX display with accuracy that is not specified.
[2] This reading can be extended up to 19999 MAX display with accuracy that is not specified.
[3] Q value is the reciprocal of DF.
[4] Cx = Counts of displayed C value. E.g., If C = 88.88 μF then Cx = 8888.
[5] Specifications are based on measurements performed at the test sockets and on battery operation.
[6] DUT and test leads need to be properly shielded by connecting to the guard terminal, if necessary.
Capacitance (Parallel Mode), Test Frequency = 10 kHz
Capacitance (Parallel Mode), Test Frequency = 100 Hz/120 Hz
Capacitance (Parallel Mode), Test Frequency = 1 kHz
Range
50 μF
20 μF
2000 nF
200 nF
20 nF
2000 pF
200pF
Range
10 mF
1000 μF
200 μF
20 μF
2000 nF
200 nF
20 nF
Range
1 mF
200 μF
20 μF
2000 nF
200 nF
20 nF
2000 pF
Maximum Display
50.0 μF
19.999 μF
1999.9 nF
199.99 nF
19.999 nF
1999.9 pF
199.99 pF
Maximum Display
19.99 mF
1999.9 μF
199.99 μF
19.999 μF
1999.9 nF
199.99 nF
19.999 nF
Maximum Display
1.999 mF
199.99 μF
19.999 μF
1999.9 nF
199.99 nF
19.999 nF
1999.9 pF
[1]
[1]
[2]
Capacitance
3.0% + 8 (DF<0.1)
3.0% + 6 (DF<0.2)
1.5% + 5 (DF<0.5)
1.5% + 5 (DF<0.5)
1.5% + 5 (DF<0.5)
2.0% + 6 (DF<0.5)
3.0% + 8 (DF<0.1)
Capacitance
3.0% + 5 (DF<0.1)
1.0% + 5 (DF<0.1)
0.7% + 3 (DF<0.5)
0.7% + 3 (DF<0.5)
0.7% + 3 (DF<0.5)
0.7% + 5 (DF<0.5)
1.0% + 5 (DF<0.1)
Capacitance
3.0% + 5 (DF<0.1)
1.0% + 5 (DF<0.1)
0.7% + 3 (DF<0.5)
0.7% + 3 (DF<0.5)
0.7% + 3 (DF<0.5)
0.7% + 5 (DF<0.5)
1.0% + 5 (DF<0.1)
Accuracy
Accuracy
7
Accuracy
DF
12.0% + 100/Cx + 10 (DF<0.1)
5.0% + 100/Cx + 8 (DF<0.2)
1.5% + 100/Cx + 6 (DF<0.5)
1.5% + 100/Cx + 6 (DF<0.5)
1.5% + 100/Cx + 6 (DF<0.5)
3.0% + 100/Cx + 6 (DF<0.1)
5.0% + 100/Cx + 8 (DF<0.1)
DF
10% + 100/Cx + 5 (DF<0.1)
2.0% + 100/Cx + 5 (DF<0.1)
0.7% + 100/Cx + 5 (DF<0.5)
0.7% + 100/Cx + 5 (DF<0.5)
0.7% + 100/Cx + 5 (DF<0.5)
0.7% + 100/Cx + 5 (DF<0.5)
2.0% + 100/Cx + 5 (DF<0.1)
DF
10% + 100/Cx + 5 (DF<0.1)
2.0% + 100/Cx + 5 (DF<0.1)
0.7% + 100/Cx + 5 (DF<0.5)
0.7% + 100/Cx + 5 (DF<0.5)
0.7% + 100/Cx + 5 (DF<0.5)
0.7% + 100/Cx + 5 (DF<0.5)
2.0% + 100/Cx + 5 (DF<0.1)
Note
After short cal.
After short cal.
After open cal.
After open cal.
Note
After short cal.
After short cal.
After open cal.
After open cal.
Note
After short cal.
After short cal.
After open cal.