ES1-LP3 Omron, ES1-LP3 Datasheet - Page 52

no-image

ES1-LP3

Manufacturer Part Number
ES1-LP3
Description
Sensor Hardware & Accessories INFRARED SENSOR 0-500 C
Manufacturer
Omron
Datasheets

Specifications of ES1-LP3

Lead Free Status / Rohs Status
Lead free / RoHS Compliant
52
E5EZ-PRR
Protect level
Communications setting level
oapt
icpt
wtpt
keyp
l.prt
l.prt
l.prt
l.prt
u-no
sbit
l.com
l.com
l.com
l.com
l.com
off
off
bps
len
9 6
0
1
1
7
2
Operation /
adjustment protect
Initial setting /
communications
protect
Setting change protect
Automatic/ manual
key protect
Communications
Unit number
Baud rate
Data length
Stop bits
Parity
Operation/ Adjustment Protect
The following table shows the relationship between set values and the range of protection.
Initial Setting/ Communications Protect
This protect level restricts movement to the initial setting level, communications setting level, and advanced
function setting level.
Setting Change Protect
This protect level restricts key operations.
Automatic/ Manual Key Protect
This protect level restricts key operations.
Setting Communications Parameters
Operation
level
Adjustment level
0
1
2
OFF
ON
OFF
ON
Communications
Unit number
Baud rate
Data length
Stop bits
Parity
Set value
Set value
Set value
Parameters are not protected when the set value is set to 0.
The default is 0.
The default is 1.
The default is OFF.
The default is OFF.
Each parameter is enabled when the power is reset.
Match the communications specifications of the E5EZ-PRR and the host computer. If multiple devices are
connected, ensure that the communications specifications for all devices in the system (except the
Communications unit number) are the same.
Parameters
Level
PV
PV/SP
Others
Movement possible
Movement possible
Movement not possible
Settings can be changed using key operations.
Settings cannot be changed using key operations. (The protect level settings, however, can be
changed.)
Settings can be changed between automatic and manual mode using A key operations.
Settings cannot be changed between automatic and manual mode using A key operations. (The
protect level settings, however, can be changed.)
Initial setting level
Can be displayed
Can be displayed
and changed
Can be displayed
and changed
Can be displayed
and changed
Symbol
u-no
sbit
prty
bps
len
0
0 to 99
1.2, 2.4, 4.8, 9.6, 19.2
7, 8
1, 2
None, even, odd
Set (monitor) values
Movement possible
Movement possible
Movement not possible
Communications setting level
Can be displayed
Can be displayed
and changed
Can be displayed
and changed
Cannot be displayed
and moving to other
levels is not possible
1
Description
Description
Set value
1.2, 2.4, 4.8, 9.6,
19.2
none, even, odd
Can be displayed
Can be displayed
and changed
Cannot be displayed
and moving to other
levels is not possible
Cannot be displayed
and moving to other
levels is not possible
Settings
Movement possible
Movement not possible
Movement not possible
Advanced function setting level
2
1
9.6
7
2
Even
Default
Can be displayed
Can be displayed
Cannot be displayed
and moving to other
levels is not possible
Cannot be displayed
and moving to other
levels is not possible
3
None
Kbps
Length
Bit
None
Unit

Related parts for ES1-LP3