OS533E NEWPORT ELECTRONICS, OS533E Datasheet - Page 40

INFRARED THERMOMETER, -18°C TO 538°C

OS533E

Manufacturer Part Number
OS533E
Description
INFRARED THERMOMETER, -18°C TO 538°C
Manufacturer
NEWPORT ELECTRONICS
Datasheet

Specifications of OS533E

Temperature Measuring Range
-18°C To 538°C
Temperature Tester Type
Thermocouple - Type K
Accuracy %
1%
Wavelength Max
670nm
Command Description
(ASCII)
IR
T
P
S
F1 or F0
E95
H500
L20
A125
t
t5
p
D0
De
2.3.13 PC Interface Commands
You can communicate directly from the PC to the infrared thermometer. Here
are the Comm port settings and communication commands from the PC:
Baud rate: 9600
Data: 8 Bits
One Stop Bit
No Parity
All the PC commands to the infrared thermometer are case sensitive and
terminates with a carriage return (CR). You can change parameter settings
from the PC when data transmission is stopped.
Here is a typical data strings from the infrared thermometer to the PC when
the “T” command is activated:
OS534; E:95; MAX:78; MIN:65; DIF:13; AVG:72; DIS:1144; HAL:900; TC:74;
TEF:0;
LAL:20; AMB:125; PRN:5; PRNF:1; IR:73; CF:0; FF:1; LF: 0:
End
Get the current infrared temperature from the thermometer
Start sending Data stings from the thermometer to the PC
Stop sending data to the PC
Reset Min, Max, Diff, Avg temperature values on the thermometer
F1 = Set Engineering unit to °F , F0 = Set Engineering unit to °C
Set Emissivity to 0.95 (Thermometer sends “E:95” back as
confirmation)
Set High Alarm set point (HAL) to 500 (It sends “HAL:500” back as
confirmation)
Set Low Alarm set point (LAL) to 20 (It sends “LAL:20” back as
confirmation)
Set Target ambient temp (AMB) to 125(It sends “AMB:125” back as
confirmation)
Get the data transmission interval (PRN) from thermometer
Set data transmission interval (PRN) to 5 seconds. Thermometer sends
back “PRN:5” as confirmation.
Get the data transmission flag from the thermometer.
PRNF:0 means no data transmission (PRN is disabled)
PRNF:1 means data transmission (PRN is enabled)
Start to download stored data from IR thermometer memory
Erase the data from the IR thermometer memory
Using the Handheld Infrared Thermometer
2
2-25