AC3006 ROTRONIC, AC3006 Datasheet - Page 6

no-image

AC3006

Manufacturer Part Number
AC3006
Description
INTERFACE CABLE, HYGROPALM PC
Manufacturer
ROTRONIC
Datasheet

Specifications of AC3006

Rohs Compliant
NA
SCS* humidity standards
ROTRONIC humidity standards are delivered in packs of five ampoules of the
same humidity value. Every ampoule is marked with its humidity value and a
serial number. The most frequently used values are 35 and 80 %rh, which are
used for two-point calibrations. All ampoules except for the 0 %rh standard
contain an unsaturated salt solution; the 0 %rh standard consists of a highly
porous molecular sieve. An SCS certificate documenting traceability to national
standard and specifying the uncertainty of the humidity standard is enclosed with
every pack. The different national agencies for metrology recognise each others'
certificates reciprocally through the ILAC Mutual Recognition Agreement. As a
result, an instrument calibration certificate from Switzerland (SCS) is accepted
worldwide by local certification bodies.
Other values on request
Probe calibration by software
and calibration interface
ROTRONIC probes can be calibrated and adjusted via the connected instrument
with either an integrated keypad or a calibration interface to a PC running HW4
software.
* SCS: Swiss Calibration Service
Order information
Order code
EA00-SCS
EA05-SCS
EA10-SCS
EA11-SCS
EA20 SCS
EA35-SCS
EA50-SCS
EA65-SCS
EA75-SCS
EA80-SCS
EA95-SCS
Humidity value
0.5 %rh
5.0 %rh
10.0 %rh
11.3 %rh
20.0 %rh
35.0 %rh
50.0 %rh
65.0 %rh
75.3 %rh
80.0 %rh
95.0 %rh
c alibr at ion
Uncertainty at 23 ± 2°C
± 0.1 %rh
± 0.1 %rh
± 0.3 %rh
± 0.3 %rh
± 0.3 %rh
± 0.5 %rh
± 0.9 %rh
± 0.9 %rh
± 0.9 %rh
± 1.2 %rh
± 1.2 %rh
DPH 911
Reference dew point mirror for
certification of SCS* humidity standards
89