53210A-400 AGILENT TECHNOLOGIES, 53210A-400 Datasheet - Page 9

Frequency Counter With GPIB Port, 10, 0.2 Ppm

53210A-400

Manufacturer Part Number
53210A-400
Description
Frequency Counter With GPIB Port, 10, 0.2 Ppm
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of 53210A-400

Bandwidth
350MHz
Counter Measuring Functions
Frequency, Period,Totalize, Ratio, Time Interval, Rise/Fall, Width, Burst Width, Duty Cycle, Phase, PRI, PRF
Lead Free Status / RoHS Status
na
Trigger and Timebase Characteristics (nom)
1. Continuous, gap-free measurements limits the Gate Time setting to 10 µs to 1000 s in 10 µs steps.
Trigger characteristics (nom)
General
Trigger source
Trigger count
Trigger delay
Samples/trigger
External trigger input (typ)
Connector
Impedance
Level
Slope
Pulse width
Latency
External trigger rate
Damage level
Timebase characteristics (nom)
Timebase reference
Timebase adjustment
method
Timebase adjustment
Resolution
External timebase input (typ)
Impedance
Level (typ)
Lock frequencies
Lock range
Damage level
Timebase output (typ)
Impedance
Level
Signal
Damage level
300/s max
53210A
±1 ppm (±0.1 ppm for Option 010 U-OCXO timebase)
10
-10
Frequency, period: 1 µs + 3 periods
Closed-box electronic adjustment
(10
Selectable positive or negative
Internal, external, bus, manual
time interval, totalize: 100 ns
-11
9
0 s to 3600 s in 1 µs steps
Internal, external, or auto
0.5 Vrms into a 50 Ω load
1.0 Vrms into a 1 kΩ load
100 mVrms to 2.5 Vrms
10 MHz, 5 MHz, 1 MHz
for Option 010 U-OCXO timebase)
50 Ω ± 5% at 10 MHz
Rear panel BNC(f)
10 MHz sine wave
1 kΩ AC coupled
TTL compatible
1 to 1,000,000
1 to 1,000,000
<-5 V, >+10 V
> 40 ns min.
1 k/s max
53220A
7 Vrms
7 Vrms
1 kΩ
10 k/s max
53230A