FLUKE-355 Fluke, FLUKE-355 Datasheet - Page 2

MULTIMETER, DIGITAL, CLAMP, 4 DIGIT

FLUKE-355

Manufacturer Part Number
FLUKE-355
Description
MULTIMETER, DIGITAL, CLAMP, 4 DIGIT
Manufacturer
Fluke
Datasheets

Specifications of FLUKE-355

Dmm Type
Clamp
Voltage Measuring Range Dc
0V To 1000V
Voltage Measuring Range Ac
0V To 1000V
Current Measuring Range Dc
0A To 2000A
Current Measuring Range Ac
0A To 2000A
Lead Free Status / RoHS Status
na
Lead Free Status / RoHS Status
na
Specifications
Electrical specifications
Current measurement dc and ac 10 Hz to 100 Hz
Crest Factor (50/60 Hz)
*Add 2 % to error spec for CF > 2
Current measurement ac 100.1 Hz to 1 kHz
Voltage measurement (355 only) dc and ac 10 Hz to 100 Hz
600 V and 1000 V ranges have 10 % over range to 660 V and 1100 V respectively.
Voltage measurement (355 only) ac 100.1 Hz to 1 kHz
600 V and 1000 V ranges have 10 % over range to 660 V and 1100 V respectively.
Ohms measurement (355 only)
2 Fluke Corporation
Range
40 A
400 A
2000 A;
1400 ac rms
Range
40 A
400 A
2000 A;
1400 ac rms
Range
4 V
40 V
400 V
600 V ac rms
1000 V dc
Range
4 V
40 V
400 V
600 V ac rms
Range
400 W
4 kW
40 kW
400 kW
Range
40 A
400 A
2000 A; 1400 ac rms
Resolution
Resolution
Resolution
Resolution
100 mV
100 mV
100 mA
100 mA
10 mV
10 mV
10 mA
10 mA
1 mV
1 mV
1 V
1 V
1 V
1 A
1 A
Fluke 355/353 True-rms 2000 A Clamp Meters
1.5 % rdg + 15 digits
3.5 % rdg + 15 digits
Crest Factor*
2 @ 33 A, 2.4 @ 27 A
2 @ 330 A, 2.4 @ 270 A
2 @ 1000 A, 2.4 @ 833 A
1.5 % rdg + 5 digits
1.5 % rdg + 5 digits
3.5 % rdg + 5 digits
3.5 % rdg + 5 digits
Accuracy > 10 A
1 % rdg + 10 digits
3 % rdg + 10 digits
Accuracy, A
1 % rdg + 5 digits
1 % rdg + 5 digits
1 % rdg + 5 digits
1 % rdg + 5 digits
3 % rdg + 5 digits
3 % rdg + 5 digits
3 % rdg + 5 digits
Accuracy
Accuracy
Resolution
Trigger Level for
Trigger Level for
100 W
0.1 W
10 W
1 W
Inrush
0.50 A
Inrush
0.50 A
5.0 A
5.0 A
5 A
5 A
Trigger Level for Hz
Trigger Level for Hz
Filter OFF
Filter OFF
0.050 V
0.050 V
0.25 V
0.25 V
N/A
6 V
6 V
6 V
6 V
Trigger Level
Trigger Level
Filter OFF
Filter OFF
for Hz
2.50 A
2.50 A
for Hz
2.5 A
2.5 A
8 A
8 A
1.5 % + 5 digits
1.5 % + 5 digits
1.5 % + 5 digits
1.5 % + 5 digits
Trigger Level for Hz
Trigger Level for Hz
Accuracy
Filter ON
Filter ON
0.050 V
0.050 V
Trigger Level
Trigger Level
0.25 V
0.25 V
N/A
6 V
6 V
6 V
6 V
Filter ON
Filter ON
for Hz
0.50 A
for Hz
0.50 A
2.5 A
2.5 A
8 A
8 A