72-7730 CALIBRATED TENMA, 72-7730 CALIBRATED Datasheet - Page 43

MULTIMETER DIGITAL HANDHELD, 4-1/2 DIGIT

72-7730 CALIBRATED

Manufacturer Part Number
72-7730 CALIBRATED
Description
MULTIMETER DIGITAL HANDHELD, 4-1/2 DIGIT
Manufacturer
TENMA
Datasheet

Specifications of 72-7730 CALIBRATED

Dmm Type
Handheld
Voltage Measuring Range Dc
200mV - 1000V
Voltage Measuring Range Ac
2V - 1000V
Current Measuring Range Dc
200µA To 10A
Current Measuring Range Ac
200µA To 10A
Lead Free Status / RoHS Status
na
2.
3.
Note
l
l
40
The LCD displays OL indicating the tested capacitor
is shorted or it exceeds the maximum range.
Capacitors larger than 400¦ÌF take longer time. The
analogue bar graph shows the time left before
finishing the measurement.
Set the rotary switch to
the Meter may display a fixed reading which is a
internal distributed capacitor value. For testing less
than 10nF capacitor, the tested value must subtract
the internal distributed capacitor value to maintain
the accuracy.
To improve the measurement accuracy of small value
capacitors (less than 10nF), press REL
test leads open to subtract the residual capacitance
of the Meter and leads.
It is recommended to use test clip to carry out
measurement to reduce the effect of internal
distributed capacitor.
measurement mode,
with the
l When capacitance measurement has been
completed, disconnect the connection between the
testing leads and the circuit under test and remove
the test leads away from the input terminals of the
Meter.
Model 72-7730: OPERATING MANUAL