FLUKE-289 Fluke Electronics, FLUKE-289 Datasheet - Page 80

DMM TRUE RMS 50000CNT W/BCKLGHT

FLUKE-289

Manufacturer Part Number
FLUKE-289
Description
DMM TRUE RMS 50000CNT W/BCKLGHT
Manufacturer
Fluke Electronics
Series
28xr
Type
Digital (DMM)r
Datasheet

Specifications of FLUKE-289

Includes
Battery, Clips, Test Leads
Style
Handheld
Display Digits
4.5
Display Type
LCD, Bar Graph
Display Count
50000
Function
Voltage, Current, Resistance, Capacitance, Temperature, Frequency
Functions, Extra
Continuity, dB
Features
Data Logging, Hold, Memory, Min/Max/Ave
Ranging
Auto/Manual
Response
True RMS
Lead Free Status / RoHS Status
Not applicable / RoHS non-compliant
Other names
289
614-1062
287/289
Users Manual
DC Voltage Specification
70
DC mV
DC V
L
[1]
[2]
[3]
[4]
[5]
Function
[1]
[1]
Add 20 counts in dual display ac over dc, dc over ac or ac+dc.
AC+DC ranges are specified from 2 % to 140 % of range, except 1000 V is specified from 2 % to 100 % of range.
When using the relative mode (REL Q) to compensate for offsets.
Add 4 counts/10 mV AC in dual display ac over dc, dc over ac or ac + dc
Add 10 counts/100 mV AC in dual display ac over dc, dc over ac or ac + dc.
50 mV
500 mV
5 V
50 V
500 V
1000 V
1000 V
Range
[3]
0.001 mV
0.01 mV
0.0001 V
0.001 V
0.01 V
0.1 V
0.1 V
Resolution
0.05 % + 20
0.025 % + 2
0.025 % + 2
0.025 % + 2
0.03 % + 2
0.03 % + 2
1 % + 20
DC
[2]
[4]
[5]
20 to 45 Hz
Not Spec’d
2 % + 80
AC over DC, DC over AC, AC + DC
45 Hz to 1 kHz
Not Spec’d
0.5 % + 80
Accuracy
1 to 20 kHz
1.5 % + 40
1.5 % + 40
Not Spec’d
Not Spec’d
Not Spec’d
1.5 % + 40
1.5 % + 40
[2]
20 to 35 kHz
Not Spec’d
Not Spec’d
Not Spec’d
5 % + 40
5 % + 40
5 % + 40
5 % + 40