FLUKE-289 Fluke Electronics, FLUKE-289 Datasheet - Page 44

DMM TRUE RMS 50000CNT W/BCKLGHT

FLUKE-289

Manufacturer Part Number
FLUKE-289
Description
DMM TRUE RMS 50000CNT W/BCKLGHT
Manufacturer
Fluke Electronics
Series
28xr
Type
Digital (DMM)r
Datasheet

Specifications of FLUKE-289

Includes
Battery, Clips, Test Leads
Style
Handheld
Display Digits
4.5
Display Type
LCD, Bar Graph
Display Count
50000
Function
Voltage, Current, Resistance, Capacitance, Temperature, Frequency
Functions, Extra
Continuity, dB
Features
Data Logging, Hold, Memory, Min/Max/Ave
Ranging
Auto/Manual
Response
True RMS
Lead Free Status / RoHS Status
Not applicable / RoHS non-compliant
Other names
289
614-1062
287/289
Users Manual
Using Conductance for High Resistance Tests
Conductance, the inverse of resistance, is the ability of a circuit
to pass current. High values of conductance correspond to low
values of resistance.
The unit of conductance is the Siemens (S). The meter’s 50 nS
range measures conductance in nanosiemens (1 nS =
0.000000001 Siemens). Because such small amounts of
conductance correspond to extremely high resistance, the nS
range is used to measure the resistance of components up to
100,000 M , or 100,000,000,000
To measure conductance, position the rotary switch to
set up the Meter as shown in Figure 18. Move the menu selector
to the menu item labeled Ohms,nS,j and press the softkey
labeled nS.
There is normally a residual conductance reading with the test
leads open. To ensure accurate readings, press the softkey
labeled Menu. Move the menu selector to the menu item labeled
REL and press the softkey labeled REL to subtract the residual
value with the test leads open.
34
R is disabled when the Meter is measuring
conductance.
Note
(1 nS = 1,000 M ).
S
and
0
Menu
Figure 18. Conductance Measurement
8:10pm
26.55
10
Save
20
30
40
nS
Manual Range
06/13/07
Setup
50 nS
est14.eps