PAXS0000 Red Lion Controls, PAXS0000 Datasheet - Page 8

Strain Gauge Meter

PAXS0000

Manufacturer Part Number
PAXS0000
Description
Strain Gauge Meter
Manufacturer
Red Lion Controls
Type
Strain Gager
Datasheet

Specifications of PAXS0000

No. Of Digits / Alpha
5
Meter Function
DC Millivoltmeter
Meter Range
± 24mV To ± 240mV
Digit Height
14.2mm
Power Consumption
15VA
Operating Temperature Range
0°C To +50°C
Supply Voltage Ac, Min
85V
Accuracy
0.07% of Reading +4 µV/+40 µV
Brand/series
PAX
Common Mode Rejection Ratio
80
Connection Type
2-Wire (Single-Ended), 4-Wire Bridge (Differential)
Cut Out, Panel
1⁄8 DIN
Dimensions
4.2" L × 3.8" W × 1.95" H
Display Digit Height
0.56 "
Display Resolution
1 μV/10 μV
Display Type
LED
Function
Strain Gage
Humidity
0 to 85% (Max.)
Impedance
100 Megohms
Input Type
Strain Gage
Length, Stripping
0.3 in.
Memory Type
Nonvolatile EEPROM
Number Of Digits
5
Primary Type
Electronic
Range, Measurement
±24/±240 mVDC
Resolution
16 bit
Special Features
Programmable Function Keys
Standards
UL Listed, CE Marked
Temperature Coefficient
20 ppm⁄°C (Max.)
Temperature, Operating
0 to +50 °C
Termination
Cage Clamp
Torque
4.5 inch-lbs (Max.)
Voltage, Input
85 to 250 VAC
Voltage, Supply
85 to 250 VAC
Character Size
0.56"
Rohs Compliant
Yes
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
prior to applying power. The following Jumper Selection Figures show an
enlargement of the jumper area.
squeezing and pulling back on the side rear finger tabs. This should lower the
latch below the case slot (which is located just in front of the finger tabs). It is
recommended to release the latch on one side, then start the other side latch.
Input Range Jumper
in programming must match the jumper setting. Select a range that is high enough
to accommodate the maximum input to avoid overloads. The selection is different
for each meter. See the Jumper Selection Figure for appropriate meter.
Excitation Output Jumper
for the application. If excitation is not being used, it is not necessary to check or
move this jumper.
Input Range Jumper
1.0 I
2.0 S
PAXD Jumper Selection
Installation
unit is intended to be mounted into an enclosed panel. Prepare the panel cutout
to the dimensions shown. Remove the panel latch from the unit. Slide the panel
The meter can have up to four jumpers that must be checked and / or changed
To access the jumpers, remove the meter base from the case by firmly
This jumper is used to select the proper input range. The input range selected
If your meter has excitation, this jumper is used to select the excitation range
One jumper is used for voltage/ohms or current input ranges. Select the proper input range high
enough to avoid input signal overload. Only one jumper is allowed in this area. Do not have a jumper
in both the voltage and current ranges at the same time. Avoid placing the jumper across two ranges.
The PAX meets NEMA 4X/IP65 requirements when properly installed. The
NSTALLING THE
ETTING THE
JUMPER SELECTIONS
The
gasket over the rear of the unit to the
back of the bezel. The unit should
indicates factory setting.
be installed fully assembled.
Insert the unit into the panel
cutout.
J
UMPERS
M
ETER
8
Installation Environment
operating temperature and provides good air circulation. Placing the unit near
devices that generate excessive heat should be avoided.
Do NOT use solvents. Continuous exposure to direct sunlight may accelerate
the aging process of the bezel.
keypad of the unit.
User Input Logic Jumper
not used, it is not necessary to check or move this jumper.
PAXH:
Signal Jumper
is installed. For voltage signals, remove the jumper from the board. (For 2 V
inputs, this removed jumper can be used in the “2 V only” location.)
Couple Jumper
removed from the board. If DC couple is used, then the jumper is installed.
so that the tabs of the panel latch engage in the slots on the case. The panel
latch should be engaged in the farthest forward slot possible. To achieve a
proper seal, tighten the latch screws evenly until the unit is snug in the panel
(Torque to approximately 7 in-lbs [79N-cm]). Do not over-tighten the screws.
While holding the unit in place, push the panel latch over the rear of the unit
The unit should be installed in a location that does not exceed the maximum
The bezel should be cleaned only with a soft cloth and neutral soap product.
Do not use tools of any kind (screwdrivers, pens, pencils, etc.) to operate the
This jumper selects the logic state of all the user inputs. If the user inputs are
This jumper is used to select the signal type. For current signals, the jumper
This jumper is used for AC / DC couple. If AC couple, then the jumper is
Main
Circuit
Board
VOLT/
OHM
JUMPER
LOCATION
PANEL CUT-OUT
CURRENT
JUMPER
LOCATION
USER INPUT
EXCITATION