PWS4602 Tektronix, PWS4602 Datasheet - Page 4

Bench Top Power Supplies Programmable DC Pwr Sup 0-60 Vlts 2 Amp

PWS4602

Manufacturer Part Number
PWS4602
Description
Bench Top Power Supplies Programmable DC Pwr Sup 0-60 Vlts 2 Amp
Manufacturer
Tektronix
Type
Power Supplyr
Datasheet

Specifications of PWS4602

Output Current (channel 1)
0 A to 2.5 A
Output Voltage (channel 1)
0 V to 60 V
Output Voltage (1)
0 V to 60 V
Output Current (1)
0 A to 2.5 A
Accuracy
0.03% Basic Voltage, 0.05% Basic Current
Display
Vaccum Fluorescent with Large Digits
Keypad Type
Numeric
Warranty
3-Year
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Data Sheet
Rear-panel Connections
Power Source
Physical Characteristics
4
Characteristic
Communications
Output, Sense, Status,
and Control
Characteristic
110 V AC Setting
220 V AC Setting
Frequency
Power Consumption
Dimension
Height
Width
Depth
Weight
PWS4205
PWS4305
PWS4323
PWS4602
PWS4721
Output channel
Remote sense lines
Control input
Status output
PWS4205/4323/4721 250 VA
PWS4305/4602
www.tektronix.com
Description
USB Device Port, Type B connector, USBTMC
compatible
GPIB, optional (requires TEK-USB-488 GPIB to USB
adapter)
Removable screw terminal block carries the following
signals:
Duplicates the front-panel outputs
Connection for remote sense
Multifunction TTL input which can function as a trigger
input, output control line, or digital input
Multifunction TTL output which can function as a fault
indication, or digital output
Description
99 V
198 V
47 Hz to 63 Hz
350 VA
RMS
RMS
to 132 V
to 264 V
mm
355
214
kg
9.0
9.6
9.0
9.6
9.0
88
RMS
RMS
3.47
8.45
13.9
19.8
21.2
19.8
21.2
19.8
in.
lb.
Environmental and Safety
Warranty
Three years.
Characteristic
Temperature
Relative Humidity (Noncondensing)
Altitude
Electromagnetic
Compatibility
Safety
Operating
Storage
Operating
Storage
Operating
Storage
Description
0 °C to +40 °C
–20 °C to +70 °C
5% to 95% relative humidity at up to +40 °C
5% to 95% relative humidity at up to +40 °C
5% to 60% RH above +40 °C up to +70 °C
Up to 2,000 m
Up to 4,000 m
European Union: EN 55011, Class A; IEC 61000-3-2;
IEC 61000-3-3, IEC 61000-4-2, IEC 61000-4-3,
IEC 61000-4-4, IEC 61000-4-5, IEC 61000-4-6, IEC
61000-4-8, IEC 61000-4-11
USA: FCC, CFR Title 47, Part 15, Subpart B, Class A
Australia: EMC Framework, demonstrated per Emission
Standard AS/NZS 2064 (Industrial, Scientific, and
Medical Equipment)
European Union: Low Voltage directive 2006/95/EC;
EN61010-1 2001
USA: Nationally recognized testing laboratory listing
UL61010-1-2004
Canada: CAN/CSA C22.2 No. 61010-1 2004

Related parts for PWS4602