FLUKE-54-2 60HZ Fluke Electronics, FLUKE-54-2 60HZ Datasheet - Page 19

THERMTR CONTCT DUAL INP W/RECRD

FLUKE-54-2 60HZ

Manufacturer Part Number
FLUKE-54-2 60HZ
Description
THERMTR CONTCT DUAL INP W/RECRD
Manufacturer
Fluke Electronics
Series
50 IIr
Type
Handheldr
Datasheets

Specifications of FLUKE-54-2 60HZ

Temperature Range
-418 ~ 3212°F (-250 ~ 1767°C)
Temperature Display
C°/F°/K°
Display Type
LCD
Input Type
Thermocouples (2)
Features
Backlight, Hold, Memory, MIN/MAX/Avg, Sleep Mode
Probe Type
J, K, T, E, N, R, S
Battery Type
AA (3)
Includes
Holster, Software, Thermocouple - K-Type Bead Probe (2)
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Other names
54-2
54-2 60HZ
614-1036
674697
Using the Offset to Adjust for Probe Errors
Use the offset option in Setup to adjust the thermometer’s
readings to compensate for the errors of a specific
thermocouple.
1.
2.
3.
4.
Plug the thermocouple into the input connector.
Place the thermocouple in a known, stable
temperature environment (such as an ice bath or a
dry well calibrator).
Allow the readings to stabilize.
In Setup change the offset until the primary display
reading matches the calibration temperature. (See
"Changing Setup Options.")
Using Memory
During a logging session the thermometer stores logged
readings in its memory.
At the end of the logging session you can view the logged
readings on the display.
You can also transfer the logged readings to a PC running
FlukeView Forms software. (See "Communicating with a
PC.")
FlukeView Forms displays the logged readings on an
online form, which you can print or store for later use.
53 & 54 Series II
Using Memory
15