I800 Fluke Electronics, I800 Datasheet - Page 3

no-image

I800

Manufacturer Part Number
I800
Description
CURRENT CLAMP PROBE AC
Manufacturer
Fluke Electronics
Type
AC Clampr
Datasheet

Specifications of I800

Style
Pincer
Ac Current Range
100mA ~ 800A rms
Bandwidth
10kHz
Cable Length
62.992" (1600.00mm)
Max Conductor Size
2.126" (54.00mm)
Termination
Banana Plug, Double, Sheathed
Ratings
CAT III 600V
For Use With
614-1062 - DMM TRUE RMS 50000CNT W/BCKLGHT
Lead Free Status / RoHS Status
Lead free by exemption / RoHS compliant by exemption
Other names
2840283
SIAC
SIAC
provides protection against short circuits between adjacent
primary conductors when clamped around conductors. This
mechanism ensures that the magnetic circuit separates at the
bottom of the jaw prior to the jaw opening at the top.
A detent action produced by SIAC™ can be felt when opening
the jaws. This detent action should always be present in use.
If the jaws open smoothly, service is indicated and the Clamp
should be inspected at the nearest Fluke Service Center. See
“Contacting Fluke”.
Operating Instructions
To avoid electric shock or personal injury when
using the Clamp:
The arrow on the labels shows the direction of the
current flow form the source to the lead. In order to have
a correct phase relationship between the primary
current and the clamp output, care must be taken
regarding the direction of current flow.
1.
2.
3.
TM
(Security Internal Anti Short Circuit), patent applied for,
Ensure that your fingers are behind the tactile
barrier as shown in Figure 1.
Do not use the Clamp if any part, including the
lead and connector(s), appears to be damaged
or if a malfunction of the instrument is
suspected.
Clamp around the current carrying conductor
ensuring good contact between the jaw faces.
Center the cable between the Clamp jaws.
Connect the Clamp output connectors between the
common and the appropriate current input jack of a
suitable meter, then switch the meter on.
Select a 20 mA ac or greater current range.
Clamp the probe around the current-carrying conductor
to be measured (see Figure 1). Make sure the probe
jaws are tightly closed around the conductor.
(Security Internal Anti Short Circuit)
WX
Note
Warning