EVB90601 Configuration Board
TEMPERATURE RANGES, UNITS CHANGING:
The temperature ranges that the IR modules are set for, must always be entered manually into the PC
software. There is no way at this time for the PC software to know what the configured temperature
ranges are inside the module.
This, along with temperature units can be changed manually like this:
Click on “Configuration”, then “Preferences”, then “Miscellaneous”. Then select
“Temperature Units” as shown here:
Select deg F or K (deg C is default).
Click on “Configuration”, then “Settings” on the main menu. Here you have to have to enter the calibrated
range of the module you are using.
The screen is shown here:
(You must always make sure you have set the correct “Settings” for the temperature ranges. There is no
means for temperature range information to be stored in the MLX90601 module EEPROM. This
temperature range information must be inserted manually, external to the module. This means if you
change any modules, you must record the temperature ranges that you have set the module to.)
RE-CALIBRATION USING THE EVB90601