CDB5530U Cirrus Logic Inc, CDB5530U Datasheet - Page 10

no-image

CDB5530U

Manufacturer Part Number
CDB5530U
Description
BOARD EVAL FOR CS5530
Manufacturer
Cirrus Logic Inc
Type
A/Dr
Datasheets

Specifications of CDB5530U

Number Of Adc's
1
Number Of Bits
24
Sampling Rate (per Second)
3.84k
Data Interface
Serial
Inputs Per Adc
1 Differential
Input Range
±2.5 V
Power (typ) @ Conditions
35mW @ 5 V
Voltage Supply Source
Analog and Digital, Dual ±
Operating Temperature
-40°C ~ 85°C
Utilized Ic / Part
CS5530
Product
Data Conversion Development Tools
Resolution
24 bit
Maximum Clock Frequency
4 MHz
Interface Type
USB
Supply Voltage (max)
5 V
Supply Voltage (min)
3.3 V
For Use With/related Products
CS5530
Lead Free Status / RoHS Status
Contains lead / RoHS non-compliant
Lead Free Status / RoHS Status
Lead free / RoHS Compliant, Contains lead / RoHS non-compliant
Other names
598-1158
CDB5530U
2.2.1.2
Window
After the user selects the appropriate setup, the Window option becomes available in the Start-Up win-
dow. The Window pull-down has three options: Start-Up Window, Setup Window, and Data Collection
Window.
2.2.2
Setup Window
The Setup window allows the user to access the internal registers of the CS5530. See Figure 5.
Figure 5. CDB5530U Setup Window
In the Setup window, all of the ADC’s registers are displayed as hexadecimal values. The Configuration
register’s settings are also defined in this window.
Update Registers: The Update Registers button will read all the register values from the CS5530 and
update the hexadecimal values displayed on the screen. It is a good idea to press the Update Registers
button when entering the Setup window, or after modifying any registers to reflect the current status of the
part.
Reset Serial Port: When this button is pressed, the software will send the synchronization sequence (dis-
cussed in the CS5530 data sheet) to the ADC. This sequence brings the CS5530’s serial port back to the
command mode, but does not reset any of the registers in the part.
Reset Part: When this button is selected, a ‘1’ is written to the RS bit in the Configuration register to reset
the part. After a reset, the registers will all return to their default values as discussed in the CS5530 data
sheet.
Read Data: This button will instruct the ADC to begin performing repeated single (fully-settled) conver-
sions on the setup channel specified by the Setup Register box. The results of the most recent conversion
10
DS742DB1