GSM25A SL Power Electronics Manufacture of Condor/Ault Brands, GSM25A Datasheet - Page 2

no-image

GSM25A

Manufacturer Part Number
GSM25A
Description
PWR SUP SWITCHER 25W TRIPLE MED
Manufacturer
SL Power Electronics Manufacture of Condor/Ault Brands
Series
GSM25r
Datasheet

Specifications of GSM25A

Rohs Status
RoHS non-compliant
Voltage - Output
5.1V, ±12V
Number Of Outputs
3
Power (watts)
25W
Applications
Medical
Power Supply Type
Switching (Open Frame)
Voltage - Input
90 ~ 264VAC
Mounting Type
Chassis Mount
1st Output
5.1 VDC @ 2.5A
2nd Output
12 VDC @ 1.5A
3rd Output
-12 VDC @ 200mA
Size / Dimension
4.25" L x 2.5" W x 0.86" H (108mm x 63.5mm x 21.8mm)
Power (watts) - Rated
25W
Operating Temperature
0°C ~ 50°C
Efficiency
70%
Approvals
CE, CSA, EN, IEC, TUV, UL
Load Regulation
1%, 4%
4th Output
-
SL Power Electronics Corp. 6050 King Drive, Bldg. A, Ventura, CA 93003, USA. Phone:(805) 486 4565 Fax:(805) 487 8911 Email: info@slpower.com Rev. 4/22/08.
Data Sheet © 2008 SL Power Electronics Corp. The information and specifications contained in this data sheet are believed to be correct at time of publication.
However, SL Power Electronics accepts no responsibility for consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without
notice.
GSC25/GSM25 MECHANICAL SPECIFICATIONS
ENVIRONMENTAL SPECIFICATIONS
Temperature (A)
Humidity (A)
Shock (B)
Altitude
Vibration (C)
Commercial
A. To maintain proper regulation on output 2, +5.1 V current must be at least 1⁄4 and not greater than 5 times V2 current. V1 must be adjusted within 1% of 5.1 V
to maintain full load regulation on V2.
B. Thermal foldback type current limit
C. Add “G” suffix to model number for RoHS compliant model.
GSC25A
GSC25D
GSC25B
Model
GSM25D
GSM25A
GSM25B
Medical
Model
Output No.
OPERATING
See individual specs.
0 to 95% RH
20 g
-500 to 10,000 ft
1.5 g
1
2
3
1
2
3
1
2
3
pk
rms’
0.003 g
2
Output
/Hz
+ 5.1 V
+12 V
+15 V
- 15 V
+5.1 V
- 12 V
+24 V
+ 5 V
-12 V
NON-OPERATING
-40 to +85°C
0 to 95% RH
40 g
-500 to 40,000 ft
5 g
rms’
pk
0.026 g
Current
2.5 A
1.5 A
0.2 A
2.5 A
1.5 A
0.2 A
2.5 A
1.0 A
0.2 A
2
/Hz
Regulation
A. Units should be allowed to warm up/operate under
non-condensing conditions before application of power.
B. Shock testing—half-sinusoidal, 10 ± 3 ms duration, ± direction,
3 orthogonal axes, total 6 shocks.
C. Random vibration—10 to 2000Hz, 6dB/octave roll-off from 350
to 2000Hz, 3 orthogonal axes. Tested for 10 min./axis operating and
1 hr./axis non-operating.
Load
1%
4%
1%
1%
4%
1%
1%
4%
1%
Initial Setpoint
Tolerance
1%
3%
4%
1%
3%
4%
1%
3%
4%
6.2 ± 0.6 V
6.2 ± 0.6 V
6.2 ± 0.6 V
Setting
OVP
Ripple and
Noise
1%
1%
1%
1%
1%
1%
1%
1%
1%
Notes
A
A
B
A
A
B
A
A
B

Related parts for GSM25A