MINT1022A1205I01 SL Power Electronics Manufacture of Condor/Ault Brands, MINT1022A1205I01 Datasheet

PWR SPLY MED SWITCHER 22W 12V

MINT1022A1205I01

Manufacturer Part Number
MINT1022A1205I01
Description
PWR SPLY MED SWITCHER 22W 12V
Manufacturer
SL Power Electronics Manufacture of Condor/Ault Brands
Series
MINT1022r
Datasheet

Specifications of MINT1022A1205I01

Voltage - Output
12V
Number Of Outputs
1
Power (watts)
22W
Applications
Medical
Power Supply Type
Switching (Open Frame)
Voltage - Input
90 ~ 264VAC
Mounting Type
Chassis Mount
1st Output
12 VDC @ 1.8A
Size / Dimension
3.5" L x 2" W x 0.67" H (88.9mm x 50.8mm x 17mm)
Power (watts) - Rated
22W
Operating Temperature
0°C ~ 50°C
Efficiency
80%
Approvals
CE, CSA, EN, UL
Line Regulation
1.5%
Load Regulation
1.5%
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
3rd Output
-
2nd Output
-
4th Output
-
Other names
271-2640
Output Power
Minimum Load
Total Regulation
Rise Time
Transient Response
Temperature Coefficient
Overvoltage Protection
Voltage Adjustability
Voltage
Frequency
Input Surge Current
Earth leakage current 264V, 60Hz
Input Current
Input Fuse
• 90-264 VAC Input
• Overvoltage and Short Circuit Protection
• Compact Size, Low Profile 2.0” x 3.5” x 0.67”
• Approved to UL, CSA and EN Standards
• CE Compliant (LVD, EMC, RoHS)
• Low Standby Power Loss (<0.5 W)
SL Power Electronics Corp.
MINT1022
22 Watt Single Output Series
Medical - Universal Input Power Supply
International Safety Standard Approvals
Specifications
Output Specifications
Input Specifications
Natural Convection
(Line and Load)
At turn-on
50% step load, 0.2A/μs
Universal Input
Thermistor limited
120VAC
230VAC
F1
6050 King Drive
Ventura, CA 93003
2 2 Watts continuous,
24 W peak for 60 sec-
onds, 10% duty cycle
Not required
1.5%, see table
100 ms max.
4 % max. dev., 1.5 msec
max. recovery to 0.5%
+/-0.03%/° C typical
Standard - See chart
F actory set with fixed
resistors to maximize
reliability
9 0-240VAC (Nominal
100-240VAC)
47-63Hz
32A max.
1 25 μA max. (200 μA
single fault)
0.45A
0.25A
2.0 A
All specifications are typical at nominal input, full load at 25°C unless otherwise stated
Phone:805.486.4565
Thermal Performance Operating ambient
Relative Humidity
Maximum Altitude
Vibration
Shock
Hold-up Time
Efficiency
Switching Frequency Fixed
Safety Approvals
Maximum Weight
MTBF
Overload and short
circuit protection
Fax:858.712.2040
Environmental Specifications
General Specifications
Email:info@slpower.com
120VAC, 60Hz
120VAC, 22W output
Medical
25° C, ground benign, 300,000 hours
Telcordia
Fully protected
(see ratings chart)
Non-operating
0-50° C
50-70° C ambient
Non-condensing
Operating
/non-operating
5Hz-500Hz
Per MIL-STD-810E
www.slpower.com
0-70° C
-40 to +85° C
Full load. See table.
derate to 50%
0-95% RH
10,000 ft./40,000 ft. max.
2.5 g rms
516.4 part IV
15ms min. at 22W output
70-80% typical depending
on output voltage
1 5 to 100kHz, varies with line
and load
U L/EN/IEC 60601-1 CSA22.2
No.601
100 Grams (0.20 lbs)
Cycling type power limit

Related parts for MINT1022A1205I01

MINT1022A1205I01 Summary of contents

Page 1

MINT1022 22 Watt Single Output Series Medical - Universal Input Power Supply • 90-264 VAC Input • Overvoltage and Short Circuit Protection • Compact Size, Low Profile 2.0” x 3.5” x 0.67” • Approved to UL, CSA and EN Standards ...

Page 2

... MINT1022 22 Watt Single Output Series Medical - Universal Input Power Supply Medical Voltage Model Output MINT1022A0505I01 5V MINT1022A1205I01 12V MINT1022A2405I01 24V Notes 1. Other output voltages available by special order 2. Regulation from initial setpoint for 50% load change from 50 100% 3. Peak output rating is 24 Watts for 1 minute 4. 20MHz Bandwidth scope, differential mode. Measured with scope probe directly across output terminals of the power supply with load terminated with 0.1 μF 5. Heatsink temperatures should not be allowed to exceed 90° Installation data is online at www.slpower.com Condor Part Number Key ...