GLM65-28 SL Power Electronics Manufacture of Condor/Ault Brands, GLM65-28 Datasheet - Page 2

PWR SPLY MEDICAL 65W 28V

GLM65-28

Manufacturer Part Number
GLM65-28
Description
PWR SPLY MEDICAL 65W 28V
Manufacturer
SL Power Electronics Manufacture of Condor/Ault Brands
Series
GLM65r
Datasheet

Specifications of GLM65-28

Rohs Status
RoHS non-compliant
Voltage - Output
28V
Number Of Outputs
1
Power (watts)
65W
Applications
Medical
Power Supply Type
Switching (Open Frame)
Voltage - Input
85 ~ 264VAC
Mounting Type
Chassis Mount
1st Output
28 VDC @ 2.3A
Size / Dimension
5" L x 3" W x 1.06" H (127mm x 76.2mm x 26.9mm)
Power (watts) - Rated
65W
Operating Temperature
0°C ~ 50°C
Efficiency
80%
Approvals
CE, CSA, EN, UL
Line Regulation
2%
Load Regulation
2%
3rd Output
-
2nd Output
-
4th Output
-
Other names
271-2162

Available stocks

Company
Part Number
Manufacturer
Quantity
Price
Part Number:
GLM65-28G
Manufacturer:
AD
Quantity:
101
SL Power Electronics Corp., 6050 King Drive, Bldg. A, Ventura, CA 93003, USA. Phone:(805) 486 4565 Fax:(805) 487 8911 www.slpower.com Rev. 01/07.
Data Sheet © 2007 SL Power Electronics, Inc. The information and specifications contained in this data sheet are believed to be correct at time of publication.
However, SL Power accepts no responsibility for consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without notice.
GLC65/GLM65 - SINGLE OUTPUT - MECHANICAL SPECIFICATIONS
* 9 A convection, 11 A with fan cooling
Commercial
Model
GLC65-5
GLC65-12
GLC65-15
GLC65-18
GLC65-20
GLC65-24
GLC65-28
GLC65-48
ENVIRONMENTAL SPECIFICATIONS
Temperature (A)
Humidity (A)
Shock (B)
Altitude
Vibration (C)
Note:
Medical
Model
GLM65-5
GLM65-12
GLM65-15
GLM65-18
GLM65-20
GLM65-24
GLM65-28
GLM65-48
OPERATING
0 to 50°
0 to 95% RH
20 g
-500 to 10,000 ft
1.5 g
pk
rms’
Output
0.003 g
5.1 V
12 V
15 V
18 V
24 V
28 V
48 V
20V
2
/Hz
NON-OPERATING
-40 to +85°C
0 to 95% RH
40 g
-500 to 40,000 ft
5 g
rms’
pk
9/11 A *
Current
3.25 A
1.35 A
0.026 g
5.5 A
4.3 A
3.6 A
2.7 A
2.3 A
2
/Hz
Regulation
A. Units should be allowed to warm up/operate under non-condensing
conditions before application of power. Derate output current and total
output power by 2.5% per °C above 50°C.
B. Shock testing—half-sinusoidal, 10 ± 3 ms duration, ± direction, 3
orthogonal axes, total 6 shocks.
C. Random vibration—10 to 2000Hz, 6dB/octave roll-off from 350 to
2000Hz, 3 orthogonal axes. Tested for 10 min./axis operating and 1hr./
axis non-operating.
Total
2%
2%
2%
2%
2%
2%
2%
2%
V1 Adjustment
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
1.20
30.28
V1 OVP Setpoint
18.5 ± 1.5 V
21.7 ± 2.0 V
24.5 ± 2.2 V
6.2 ± 0.6 V
14 ± 1.1 V
28 ± 2.5 V
34 ± 2.8 V
55 ± 4.0 V
and Noise
Ripple
1%
1%
1%
1%
1%
1%
1%
1%

Related parts for GLM65-28