ADIS16209_08 AD [Analog Devices], ADIS16209_08 Datasheet - Page 15

no-image

ADIS16209_08

Manufacturer Part Number
ADIS16209_08
Description
High-Accuracy, Dual-Axis Digital Inclinometer and Accelerometer
Manufacturer
AD [Analog Devices]
Datasheet
Status
The STATUS register provides a series of error flags that
provide indicator functions for common system-level issues.
All of the flags clear (set to 0) after each STATUS register read
cycle. If an error condition remains, the error flag returns to 1
during the next sample cycle.
Rev. 0 | Page 15 of 16
Table 21. STATUS Bit Descriptions
Bit
15:10
9
8
7:6
5
4
3
2
1
0
Description
Not used
Alarm 2 status:
Alarm 1 status
Not used
Self-test diagnostic error flag
Not used
SPI communications failure
Flash update failed
Power supply above 3.625 V
Power supply below 2.975 V
1 = active, 0 = inactive
1 = active, 0 = inactive
1 = error condition, 0 = normal operation
1 = error condition, 0 = normal operation
1 = error condition, 0 = normal operation
1 ≥ 3.625 V, 0 ≤ 3.625 V (normal)
1 ≤ 2.975 V, 0 ≥ 2.975 V (normal)
(Default = 0x0000)
ADIS16209

Related parts for ADIS16209_08