E1412A Agilent Technologies, Inc., E1412A Datasheet - Page 4

no-image

E1412A

Manufacturer Part Number
E1412A
Description
6.5-digit High-accuracy Multimeter C-size
Manufacturer
Agilent Technologies, Inc.
Datasheet
DC Specifi cations
Specifi cations are for 1-hour warm-up at an integration time of 100 PLCs.
DC:DC Ratio:
Measurement method:
Input HI to Input LO:
Reference HI to Input LO:
Reference LO to Input LO:
Additional Error with Autozero OFF:
Following instrument warm-up at calibration temperature ± 1°C and <10 minutes.
100 mV-100 V ranges:
300 V range:
DC Operating Characteristics:
Function
DCV, DCI and Ω
DCV, DCI and Ω
DCV, DCI and Ω
DCV, DCI and Ω
DCV, DCI and Ω
*For 300 V range: use 0.003% of range for 5.5 digits and 0.030% of range for 4.5 digits.
For all ranges: add 20 µV for dc Volts, 4 µA for dc current, or 20 mΩ for resistance.
Considerations
Settling considerations:
Measurement considerations:
(continued)
Input HI-LO/Reference HI-LO
Apply “Reference HI-LO” signal to Ohms 4-Wire Sense terminals.
100 mV to 300 V
<12 V on 100 mV to 10 V ranges (autoranged)
<2 V
add 0.0002% reading + 5 µV
add 0.0006% reading
NPLC
100
10
1
0.2
0.02
Readings speeds for 60 Hz and (50 Hz) operation, Autozero OFF.
Digits
6.5
6.5
5.5
5.5
4.5
Reading settling times are affected by source impedance, cable
dielectric characteristics, and input signal changes.
Agilent recommends the use of Tefl on or other high impedance,
low-dielectric absorption wire insulation for these measurements.
Reading/s
0.6 (0.5)
6 (5)
60 (50)
300
1,000
4
Additional Noise Error
0% of range
0% of range
0.001% of range*
0.001% of range*
0.01% of range*