VR001 Mide Technology Corporation, VR001 Datasheet - Page 3

VIBRATION DATA LOGGER

VR001

Manufacturer Part Number
VR001
Description
VIBRATION DATA LOGGER
Manufacturer
Mide Technology Corporation
Series
Slam Stick™, Volture™r
Datasheets

Specifications of VR001

Featured Product
Slam Stick™ Vibration Recorder
Accessory Type
Vibration Characterization Device
Lead Free Status / Rohs Status
Lead free by exemption / RoHS compliant by exemption
For Use With/related Products
Volture™ Vibration Energy Harvesters
For Use With
V25W - PIEZOELECTRIC ENERGY HARVESTERV20W - PIEZOELECTRIC ENERGY HARVESTERV21BL - PIEZOELECTRIC ENERGY HARVESTERV21B - PIEZOELECTRIC ENERGY HARVESTERV22BL - PIEZOELECTRIC ENERGY HARVESTERV22B - PIEZOELECTRIC ENERGY HARVESTER
PERFORMANCE PLOTS
samples which contain this amount of error. The RMS
error (or standard error) is the error value at which
68.2% of samples (one standard deviation) have an
error lower than the RMS value.
Frequency roll off data was taken by exciting the Slam
Stick at a constant amplitude (1 gee in this case) over
a range of frequencies from 10 Hz to 3200 Hz. The
measured vibration amplitude from the Slam Stick was
then plotted versus frequency. Note that the roll off is
REVISION N0. 001
REVISION DATE: 08-24-2010
severe after ~240 Hz. Different devices performed
differently after this frequency and amplitude data
should not be trusted when above this frequency.
Frequency data is not impacted by this roll-off. For
example if a vibration source was providing a 3 gee
amplitude at 500 Hz, the Slam Stick would properly
provide the relevant frequency information, however
the gee level would be inaccurate, most likely providing
a lower than 3 gee amplitude reading.
www.mide.com
3

Related parts for VR001