MSOX3012A-GSA AGILENT TECHNOLOGIES, MSOX3012A-GSA Datasheet - Page 20

OSCILLOSCOPE, 2 ANLG + 16 DGTL, 100MHZ, 4GSPS

MSOX3012A-GSA

Manufacturer Part Number
MSOX3012A-GSA
Description
OSCILLOSCOPE, 2 ANLG + 16 DGTL, 100MHZ, 4GSPS
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of MSOX3012A-GSA

Scope Type
Digital Bench
Scope Channels
2 Analog + 16 Digital
Bandwidth
100MHz
Meter Display Type
WVGA Color
Sampling Rate
4GSPS
Input Impedance
1Mohm
Rise Time
3.5ns
Rohs Compliant
NA
Oscilloscopes redefined: Breakthrough technology delivers more scope for the same budget
Performance characteristics
Trigger type selections
Characteristics
Edge
Pulse width
Runt
Setup and hold
Rise/fall time
N
Pattern
Time-qualified pattern
Video
USB
I²S (optional)
I
SPI (optional)
CAN (optional)
LIN (optional)
RS-232/422/485/UART (optional)
2
C (optional)
th
edge bust
All 3000 X-Series models
Trigger on a rising, falling, alternating or either edge of any source
Trigger on a pulse on a selected channel, whose time duration is less than a value, greater
than a value, or inside a time range
Trigger on a position runt pulse that fails to exceed a high level threshold. Trigger on a
negative runt pulse that fails to exceed a low level threshold. Trigger on either polarity runt
pulse based on two threshold settings. Runt triggering can also be time-qualified (< or >) with
a minimum time setting of 4 ns and maximum timesetting of 10 s.
Trigger and clock/data setup and/or hold time violation from < 0.0 to 10 s
Trigger on rise time or fall time edge speed violations (< or >) based on user-selectable
threshold. Time settings range from (< or >) or 2 ns to 10 s.
Trigger on the N
Trigger when a specified pattern of high, low, and don’t care levels on any combination of
analog, digital , or trigger channels is [entered | exited]. Pattern must have stabilized for a
minimum of 2 ns to qualify as a valid trigger condition.
Trigger on a multi-channel pattern whose time duration is less than a value, greater than a
value, greater than a time value with a timeout, or inside or outside of a set of time values.
• Minimum duration setting: 2 ns - 10 ns (depends on bandwidth)
• Maximum duration setting: 10 s
Video – Trigger on all lines or individual lines, odd/even or all fields from composite video, or
broadcast standards (NTSC, PAL, SECAM, PAM-M)
Trigger on start of packet, end of packet, reset complete, enter suspend, or exit suspend.
Support USB low-speed and full-speed.
Trigger on 2’s complement data of audio left channel or right channel (=, ≠, <, >, > <, < >,
increasing value, or decreasing value)
Trigger on I
with address and/or data values. Also trigger on missing acknowledge, address with no accq,
restart, EEPROM read, and 10-bit write.
Trigger on SPI (Serial Protocol Interface) data pattern during a specific framing period.
Supports positive and negative Chip Select framing as well as clock Idle framing and user-
specified number of bits per frame.
Trigger on CAN (controller area network) version 2.0A and 2.0B signals. Trigger on the start
of frame (SOF) bit (standard). Remote frame ID (RTR), data frame ID (~RTR), remote or data
frame ID, data frame ID and data, error frame, all errors, acknowledge error and overload
frame.
Trigger on LIN (Local Interconnect Network) sync break, sync frame ID, or frame ID and data.
Trigger on Rx or Tx start bit, stop bit or data content
• Maximum duration setting: 10 s
• Minimum duration setting: 2 ns- 10 ns (depends on bandwidth)
2
C (Inter-IC bus) serial protocol at a start/stop condition or user defined frame
th
edge of a burst that occurs after a specified idle time.
20