ACD-3300 IND Amprobe, ACD-3300 IND Datasheet - Page 4

Clamp Multimeters & Accessories AC 1000A INDUSTRIAL CLAMP

ACD-3300 IND

Manufacturer Part Number
ACD-3300 IND
Description
Clamp Multimeters & Accessories AC 1000A INDUSTRIAL CLAMP
Manufacturer
Amprobe
Datasheet

Specifications of ACD-3300 IND

Ranging
Auto
Frequency
40 Hz to 20 KHz
Maximum Ac Current
1000 A
True Rms
Yes
Voltage
750 VAC, 1000 VDC
Lead Free Status / Rohs Status
Lead free / RoHS Compliant
Other names
3434877
ACD-3300 IND CAT IV Industrial True RMS Clamp Meter
Technical Data – General Information
Amprobe® Test Tools
website: www.Amprobe.com
email: info@amprobe.com
Everett, WA 98203
Tel: 877-AMPROBE
Amprobe® Test Tools
Display
Polarity
Over range
Zero
Low battery indication
Auto power off
Backlight
Measurement rate
Analog bar-graph
Operating environment
Storage temperature
Temperature Coefficient
Environmental
Jaw opening capability
Power
Battery life
Dimensions
Weight
Safety
EMC
3¾ digit liquid crystal display (LCD) (6600 count) with a 66-segment analog bar-graph
Automatic, positive implied, negative polarity indication
(OL) or (-OL) is displayed
Automatic
“” is displayed when the battery voltage drops below the operating level
Approx. 30 minutes
Backlight auto-off approx. 60 sec.
2.8 times per second, nominal
28 times per second
0°C to 50°C (32°F to 122°F) at < 70% R.H.
-20°C to 60°C (-4°F to 140°F) at < 80% R.H. with battery removed from meter
0.1 × (specified accuracy) per °C. (0°C to 18°C, 28°C to 50°C)
2000m (6561.7 Feet), Indoor use
57 mm (2.0 in) conductor
Single standard 9-volt battery, NEDA 1604, JIS 006P, IEC 6F22
typically 75 hours with carbon-zinc; 150 hour with alkaline
283 x 105 x 50 mm (11 x 4 x 2.1 in.)
500 gm ( 1.10 lb.)
LVD Meets EN61010-1:2001 and EN61010-2-032:2002, CAT III 1000V, CAT IV 600V, class II and pollution
degree 2
EN 61326-1:2006 This product complies with requirements of the following European Community
Directives: 2004/108/EC (Electromagnetic Compatibility) and 2006/95/EC (Low Voltage) as amended
by 93/68/EEC (CE Marking). However, electrical noise or intense electromagnetic fields in the vicinity
of the equipment may disturb the measurement circuit. Measuring instruments will also respond to
unwanted signals that may be present within the measurement circuit. Users should exercise care and
take appropriate precautions to avoid misleading results when making measurements in the presence
of electronic interference.
Amprobe® Test Tools Europe
Amprobe Test Tools Europe
Beha-Amprobe GmbH
In den Engematten 14
79286 Glottertal, Germany
Tel.: +49 (0) 7684 8009 - 0
Optional Accessories
TL-1500 Test leads with set of alligator clips
©2009 Amprobe Test Tools. All rights reserved.
03/2009 3465178 Rev A
www.Amprobe.com
Data Sheet