S P E C I F I C A T I O N S
Measured Components
L(Inductance),C(Capacitance), R(Resistance)
Measurement Circuit configurations:
Measurements
Resistance:
Measured Parameter:
Measurement Ranges:
Inductance:
Measured Parameter:
Measurement Range:
Capacitance:
Measured Parameters:
Measurement Range:
Series and Parallel
R: 10µΩ to 100MΩ
Q: 0.0001 to 9,999
L: 0.01µH to 99,999 H
Q: 0.0001 to 9,999
C: 0.00001pF to 99,999µF
D: 0.0001 to 9,999
R: 0.0001Ω to 9,999KΩ
R+Q
L+Q
C+D and C+R
High Accuracy, Wide Range LCR Meter
with LCD Readout
0.02% Basic accuracy
Wide measurement Ranges for Capacitors, Resistors and Inductors
Menus and parameters are displayed on an easy to read 320x 240 Graphics LCD with backlight
Measurement Rates to 12 times per second
500 Test Frequencies from 12Hz to 100KHz
Open and short-circuit compensation for accurate zeroing
Binning and Pass/Fail operation
Measurement averaging from 1 to 255
Optional GPIB and handler interface
LCD displays measured values, Q/D, Absolute
Display
Display Type:
Display value:
Parameters Displayed:
and %
Test Environment
Test Frequencies:
12Hz to 100KHz
Test Voltage:
Test Voltage Accuracy:
Measurement rates:
Measurement Delay:
Ranging:
DC Bias:
Triggering:
Zeroing correction:
Averaging:
Slow 1 Meas/sec
Medium: 5 Meas/sec
Fast: 12 Meas/sec
Internal: 2V, External: 0 to 60V
Auto or Manual.
Continuous or Triggered
1 to 255 Avgs
5mV to 1.275V in 5mV increments
320x 240 Graphics LCD W/Backlight
5 digit LCR and a 4 digit Q/D measurement
500 test frequencies from
Open and short circuit
from 1ms to 99,999ms
Measured values, absolute
±(5% + 5mV) (1 + 001 x F2)
Z9218
and %
Measurement Accuracy
Basic accuracy:
(See user manual for all accuracy conditions)
Remote Operation:
General Specifications
AC Voltage input:
Frequency:
Power Consumption:
Operating Temperature:
0 to 50°C (32 to 122°F) at
Size:
Weight:
Accessories:
GPIB Handler (Optional)
5.5" (H) x 15" (W) x 18" (D)
17.6 lbs
50/60 Hz
Manual, Line Cord, Kelvin leads with bag,
± >0.02% at 23°C ±5°C (73°F ±9°F),
120 or 220 Volts
GPIB and Handler (Optional)
20Watts (Max)
CE
85% Relative Humidity

Z9218 Summary of contents

Page 1

... R: 0.0001Ω to 9,999KΩ Protek Test & Measurement 45 Smith Street, Englewood, NJ 07631 Tel: 201-818-7711 Fax: 201-760-9888 32 Website: www.protektest.com Email: sales@protektest.com and % Z9218 Display Display Type: 320x 240 Graphics LCD W/Backlight Display value: 5 digit LCR and a 4 digit Q/D measurement Parameters Displayed: ...

Page 2

... Electrolytic for an accurate reading. C/R % MEASUREMENT SCREEN SHORT CIRCUIT CALIBRATION SCREEN 45 Smith Street, Englewood, NJ 07631 Tel: 201-818-7711 Fax: 201-760-9888 Website: www.protektest.com Email: sales@protektest.com Z9218 Selects the type of component and parameter to be measured. R&Q, L&Q, C&D, C&R or Auto may be selected. The start key initiates a measurement in the triggered mode ...