30C-NIST Protek, 30C-NIST Datasheet

no-image

30C-NIST

Manufacturer Part Number
30C-NIST
Description
Digital Multimeters LOOP CALIBRATOR
Manufacturer
Protek
Datasheet

Specifications of 30C-NIST

Lead Free Status / Rohs Status
Lead free / RoHS Compliant
18
Protek Test & Measurement
45 Smith Street, Englewood, NJ 07631 Tel: 201-818-7711 Fax: 201-760-9888
Website: www.protektest.com Email: sales@protektest.com
30C
Measurements:
Voltage:
Ranges:
Best Resolution:
Accuracy:
Overload Protection:
Current:
Ranges:
Accuracy:
Best
Sourcing
Current:
Range:
through overlapping)
Accuracy:
Loop power:
Voltage
Range:
Accuracy:
TX simulator
Type:
External Excitation Voltage:
Simulation current:
obtained through overlapping)
Output current Ramp:
Range:
Ramp Times:
S P E C I F I C A T I O N S
Resolution:
2-wire transmitter simulation
4 to 20mA (Values from 0 to 24mA maybe obtained
2V and 20V
4mA to 20mA
2, 20, 50Volts
2mA and 20mA
±(0.05% + 5d)
±(0.025% + 5d)
±(0.05% + 5d)
±(0.05% + 5d)
24V
5s, 20s, 50s
0.1µA
100µV
4 to 20mA (Values from 0 to 24mA maybe
100VDC or AC peak
30C
Loop Calibrator
3V min to 50V max
24 volt internal loop supply
Auto Step and Auto ramp modes; no need for external mA meter
Simultaneous mA and % read out fo quick, easy interpretation of reading
1µA resolution for mA source, simulate and measure
25% step button for quick linearity tests
Large 4 ½ digit 20,000 count LCD display
Encoder dial for easy one handed operation
General Specifications:
Operating Temperature:
Power:
Size:
Weight:
Accessories:
and 9 Volt battery
7.0" (H) x 3.5"(W) x 1.5" (D)
(1) 9V ANSI/NEDA 1604 battery
10.5 oz
Voltage Measure Display
Loop Power Display
Test Leads, Instruction manual, carrying case
-10 to 55ºC (14 to 131ºF)
Voltage Source Display
Simulator Display
mA Source Display