Specifications of FLUKE-284

Frequency
1 mHz to 10 MHz
Equipment Type
Waveform Generator
Lead Free Status / Rohs Status
Lead free / RoHS Compliant
Other names
2423199
12
Fluke 280 and 290 Series
Inputs
Trig in
Modulation in
Sum in
Ref clock in/out
ARB clock in/out
290 Series generators can use an external signal as the arbitrary
waveform clock. The 292 and 294 also include an internal system
clock generator (in addition to the individual channel clock genera-
tors). The output of this system clock can be made available to drive
external circuitry or the input of another generator.
Frequency range
Signal range
Min. pulse width
Polarity
Input impedance
Frequency range
VCA signal range
SCM signal range
Input impedance
Frequency range
Signal range
Input impedance
Set to input
Set to output
Set to phase lock
Set to input
Set to output
Frequency range
Max. input voltage
280 Series
N/A
N/A
N/A
N/A
280 Series
DC to 8 MHz
Approximately 2 V pk-pk input for 20 V pk-pk out put
Typically 1 kΩ
280 Series
Input for an external 10MHz reference clock.
TTL/CMOS threshold level
Buffered version of the internal 10 MHz clock. Output
levels nominally 1 V and 4 V from 50 Ω
Used together with SYNC OUT on a master and
TRIG IN on a slave to synchronize (phase lock) two
separate generators
280 Series
DC to 100 kHz
Approximately 1 V pk-pk for 100 % level change at
maximum output
Approximately ± 1 V pk for maximum output.
Typically 1 kΩ
280 Series
DC to 1 MHz
Threshold nominally TTL
level; maximum input
± 10 V
50 ns, for Trigger/Gate; 50 us for Sweep mode
Selectable as high/rising edge or low/falling edge
Typically 10 kΩ
291
Input for an external Arb clock.
TTL/CMOS threshold level.
N/A
DC to 50 MHz.
+5 V, -1 V
291
DC to 30 MHz
292/294
Outputs System Clock,
logic level < 0.8 V to > 3 V
290 Series
290 Series
DC to 100 kHz
290 Series
Threshold adjustable
over ± 5 V range;
maximum input ± 10 V
292/294
DC to 16 MHz
Hold
Holds an arbitrary waveform at
its current position. A TTL low
level or switch clo sure causes the
waveform to stop at the current
position and wait until a TTL high
level or switch opening which
allows the waveform to continue.
The front panel MAN HOLD key
or remote command may also be
used to control the Hold function.
While held the front panel MAN
TRIG key or remote command
may be used to return the wave-
form to the start. The Hold input
may be enabled independently
for each channel. Input imped-
ance is10kΩ.
Channel relationships
The channels of a multi-channel
unit can be operated entirely
independently, as if they were
separate generators. The “copy”
key allows the settings of any
channel to be instantly copied to
another when required.
relationships of modulation,
summing, triggering, or phase
locking can be set up.
Multi channel sweep
Any number of channels may be
swept simultaneously. Amplitude,
Offset and Waveform can be set
independently for each channel.
For 280 Series units, the sweep
parameters will be the same
for all channels. For 290 Series
units,the sweep parameters can
be set independently for each
channel.
Alternatively, inter-channel
ARB clock out
See ARB clock In/Out within INPUTS section.
Cursor/marker out (280 Series units only)
Adjustable output pulse for use as a marker in
sweep mode or as a cursor in arbitrary wave-
form editing mode. Can be used to modulate the
Z axis of an oscilloscope or be displayed on a
second ‘scope channel.
System clock (292/294 only)
The 292/294 units incorporate an additional
frequency generator, which can be used as a
clock source for multi-channel arbitrary wave-
forms and as an auxiliary output.
The output of the system clock is available as a
1 volt pk-pk sinewave at the Auxiliary Sine Out
socket, and as a logic level squarewave at the
Ext. ARB In/Out socket.
multi-channel arbitrary waveforms, the system
clock provides an independent fixed amplitude
sine or square output which is additional to the
two or four main channel outputs.
Signal level:
Output
impedance:
Frequency
range
Frequency
resolution
When not being used as a clock source for
280
Series
N/A
N/A
Adjustable from
nominally 2 V to 14
V, normal or inverted;
ad justable width as a
cursor
600 Ω typical
280 Series
291
N/A
N/A
292/294
DC to 50 MHz
0.1 Hz
N/A
N/A
290 Series
Inter-channel modulation
The waveform from any channel may be used to Amplitude
Modulate (AM) or Suppressed Carrier Modulate (SCM) the next
channel. Alternatively any num ber of channels may be
Modulated (AM or SCM) with the signal at the MODULA TION
input socket.
Inter-channel analog summing
Waveform Summing sums the waveform from any channel into
the next channel. Alternatively any number of channels may be
summed with the signal at the SUM input socket.
Carrier frequency
Carrier waveforms
Modulation types
Modulation source
Frequency range
Internal AM depth
Internal AM resolution
Carrier suppression
(SCM)
External modulation
signal range
Carrier frequency
Carrier waveforms
Sum source
Frequency range
External signal range
Input impedance
282/284
Entire range for selected waveform
All standard and arbitrary waveforms
Internal from the previous channel. External from
SUM IN socket
DC to > 8 MHz.
Approx. 5 V pk-pk input
for 20 V pk-pk output
Typically 1 kΩ
282/284/292/294
Entire range for selected waveform
All standard and arbitrary waveforms
AM: Double sideband with carrier. SCM: Double
sideband suppressed carrier.
Internal from the previous channel. External from
Modulation input socket. The external modulation
signal may be applied to any num ber of channels
simultaneously.
DC to > 100 kHz
0 % to 105 %.
1 %
> 40 dB
VCA: Approximately 1V pk-pk for 100 % level
change at maximum output. SCM: Approximately ±
1 V pk for max. output.
Single and Multi-Channel Universal ARB Generators
292/294
DC to > 16 MHz.
Approx. 2 V pk-pk input
for 20 V pk-pk output