53181A Agilent Technologies Test Equipment, 53181A Datasheet

no-image

53181A

Manufacturer Part Number
53181A
Description
COUNTER, 225 MHZ RF COUNTER 10 DIGITS/S
Manufacturer
Agilent Technologies Test Equipment
Datasheet
• 225 MHz bandwidth
• 10- or 12-digit resolution
• GPIB interface and
• Data transfer rate of up
Agilent 53131A/132A/181A Counters
High-performance, low-cost counters simplify and
speed systems and bench frequency measurements
Data Sheet
(optional 1.5, 3, 5,
or 12.4 GHz)
with 1 s gate time
IntuiLink connectivity
software standard
to 200 fully formatted
measurements/second
Agilent Technologies
A family of universal and
RF counters to meet your needs
Agilent Technologies 53131A/
132A/181A high-performance
counters give you fast, precise
frequency measurements at
an affordable price. These
counters feature an intuitive
user interface and one-button
access to frequently used func-
tions so you can make accurate
measurements quickly and easily.

53181A Summary of contents

Page 1

... IntuiLink connectivity software standard • Data transfer rate 200 fully formatted measurements/second Agilent Technologies A family of universal and RF counters to meet your needs Agilent Technologies 53131A/ 132A/181A high-performance counters give you fast, precise frequency measurements at an affordable price. These counters feature an intuitive user interface and one-button ...

Page 2

... In addition, the 53132A offers advanced arming modes for time interval measurements. Agilent 53181A RF Counter Optimized for RF applications, the single-channel 10 digit/s 53181A measures frequency, period and peak voltage. A digit- blanking function lets you easily ...

Page 3

... Agilent IntuiLink provides easy access to the counter’s data from your PC The Agilent 53131A/132A/181A counters, capture precise frequen- cy and time measurements. IntuiLink software allows that data to be put to work easily. You work in a familiar environment at all times, using PC applications such as Microsoft Excel® or Word® ...

Page 4

... When the 53131A or 53132A are ordered with the optional rear terminals (Opt. 060), the channel 1 and 2 inputs are active on both front and rear of the counter. When the 53181A is ordered with the optional rear terminal, the channel 1 input is active on both front and rear of the counter. For this condition, specifications indicated for the rear connections also apply to the front connections ...

Page 5

... Gate Time Period 53132A t 53181A t res res res 200 ps 650 ps N Gate Time = Frequency Frequency t acc ± Time Base Error ± Gate Time 53132A t 53181A t acc acc acc 100 ps 350 ps 500 ps 1. res jitter Gate Time Number of Samples Gate Time ...

Page 6

... Channel 2 (53181A) Option 015 0. (53181A only) Option 030 0. Option 050 0 Option 124 Frequency Ratio (53131A, 53132A, 53181A) Measurement is specified over the full signal range of each input. -10 11 Results Range “Auto” Gate Time 100 ms Time Interval (53131A, 53132A) Measurement is specified over the full signal ...

Page 7

... External Arming: An edge on the External Arm Input enables the start of each measurement. Auto Arming, Timed arming modes or another edge on the External Arm Input may be used to complete the measurement Stop Trigger Error Ch1 , Ch2 (53131A and 53132A) (53181A) Ch2 Ch1 Ch2 Freq 2 Ratio : 2 (Ch1 Freq) Gate Time 1 2 ...

Page 8

... Multiply this quantity by the measurement result to yield the absolute error for that measurement. Averaging measurements will not reduce (fractional) time base error. The counters exhibit negligible sensitivity to line voltage; consequently the line voltage term may be ignored. Trigger Error External source and input amplifier noise may advance or delay the trigger points that define the beginning and end of a measurement ...

Page 9

Graph 1: Agilent 53131A/181A–Worst Case RMS Resolution 7 (Automatic or External Arming) The graphs may also be used to compute errors for Period Measurements. To find the Period error (DP), calculate the frequency of the input signal (F = 1/P) ...

Page 10

Graph 4: Agilent 53132A–Worst Case RMS Resolution 7 (Automatic or External Arming) Graph 5: Agilent 53132A–Worst Case RMS Resolution 7 (Time or Digits Arming) Graph 6: Trigger Level Timing Error (Level Setting Error and Input Hysteresis) 7. Graphs 1, 2, ...

Page 11

Measurement Statistics Available Statistics Mean, Minimum, Maximum, Standard Deviation Number of Measurements 2 to 1,000,000. Statistics may be collected on all measurements or on only those which are between the limit bands. When the Limits function is used in conjunction ...

Page 12

... Opt. 050 5 GHz RF input with type N connector 53181A) Opt. 124 12.4 GHz RF input with type N connector 53181A) Opt. 060 Rear-panel connectors* Opt. 0B0 Delete manual set Opt. A6J ANSI Z540 compliant calibration Opt. 1CM Rack mount kit (P/N 5063-9240)** ...

Page 13

Agilent Email Updates www.agilent.com/find/emailupdates Get the latest information on the products and applications you select. Agilent Direct www.agilent.com/find/quick Quickly choose and use your test equipment solutions with confidence. Agilent Open www.agilent.com/find/open Agilent Open simplifies the process of connecting and programming ...