Datasheet
Electrical specifications
DC specifications for U1271A and U1272A
1. Notes for voltage specifications:
2. Notes for resistance specifications:
Resistance
Diode Test
Voltage
Current
Function
The accuracy of the 30 to 300 mV range is specified after the Null function is used to subtract the thermal effect (by shorting the test leads).
For Z
Overload protection: 1000 Vrms for short circuits with <0.3 A current.
Maximum open voltage is <+3.3 V
Built-in buzzer beeps when the resistance measured is less than 25 Ω ± 10 Ω. The multimeter can capture intermittent measurements longer
than 1 ms.
The accuracy of the 300 Ω to 3 kΩ range is specified after the Null function is used to subtract the test lead resistance and thermal effect (by
shorting the test leads).
For the ranges of 30 MΩ and 100 MΩ, the RH is specified for <60%.
The accuracy for ranges <50 nS is specified after the Null function is used on an open test lead.
The temperature coefficient of the 100 MΩ and 300 MΩ range is 0.1 × (specified accuracy)/°C (from -20 °C to 18 °C or 28 °C to 55 °C)
LOW
[1]
[3]
measurements, autoranging is disabled and the multimeter’s range is set to 1000 volts in the manual ranging mode.
[4]
[2]
enabled, applicable for 1000 V
Z
range and resolution only.
LOW
3000 µA
100 MΩ
300 MΩ
Range
300 mV
300 mA
300 kΩ
300 µA
1000 V
30 MΩ
300 nS
30 mV
30 mA
300 Ω
30 kΩ
300 V
3 MΩ
30 Ω
Auto
30 V
3 kΩ
10 A
3 A
3 V
3 V
(low input impedance)
Resolution
0.0001 MΩ
0.0001 kΩ
0.001 MΩ
0.001 mA
0.001 mV
0.0001 V
0.001 kΩ
0.01 MΩ
0.01 MΩ
0.0001 A
0.0001 V
0.0001 V
0.01 mV
0.01 mA
0.001 V
0.001 Ω
0.01 kΩ
0.01 µA
0.001 A
0.01 nS
0.01 Ω
0.1 µA
0.01 V
0.1 V
counts of least significant digit)
U1271A
0.05 + 5
0.05 + 5
0.05 + 2
0.05 + 2
0.05 + 2
2.0 +10
0.3 +10
0.3 +10
0.2 + 5
0.2 + 5
0.2 + 5
0.5 + 5
0.6 + 5
1.2 + 5
0.2 + 5
0.2 + 5
0.2 + 5
0.2 + 5
0.5 + 5
1 + 10
Accuracy ±(% of reading +
-
-
-
-
-
8
2.0 % + 10 @ <100 MΩ
8.0 % + 10 @ >100 MΩ
U1272A
0.05 + 20
0.05 + 5
0.05 + 5
0.05 + 2
0.05 + 2
0.05 + 2
0.2 + 10
0.3 +10
0.3 +10
0.2 + 5
0.2 + 5
0.2 + 5
0.5 + 5
0.6 + 5
1.2 + 5
0.2 + 3
0.2 + 3
0.2 + 3
0.2 + 3
0.5 + 5
0.5 + 5
1 + 20
1 + 10
-
Approximately 0.1 to 0.3 mA
Approximately 1 to 2 mA
Burden voltage
Test current/
<0.04 V/ 100 Ω
93 nA//10 MΩ
93 nA//10 MΩ
93 nA//10 MΩ
93 nA//10 MΩ
93 nA//10 MΩ
<0.1 V/ 0.01 Ω
<0.3 V/ 0.01 Ω
<0.4 V/ 100 Ω
<0.08 V/ 1 Ω
<1.00 V/ 1 Ω
0.65 mA
0.65 mA
0.65 mA
6.5 mA
65 mA
-
-
-
-
-