U2004A Agilent Technologies Test Equipment, U2004A Datasheet - Page 2

no-image

U2004A

Manufacturer Part Number
U2004A
Description
USB Power Sensor; 9 kHz - 6 GHz
Manufacturer
Agilent Technologies Test Equipment
Datasheet
Features
Table 1: Low power path and high power path for Agilent U2000 Series USB power sensors
Table 2: Product specifi cations for Agilent U2000 Series USB power sensors
Sensor
U2000A, U2001A, U2002A, U2004A
U2000A
U2001A
U2002A
U2004A
Model
Frequency range from 9 kHz to 24
GHz (sensor dependent)
Perform power measurement
without a power meter
Dynamic range from
+20 dBm
Internal zeroing capability
and external calibration-free
measurements
Simplifi ed measurement setup
with built-in triggering
Frequency Range
10 MHz to 18.0 GHz
10 MHz to 6.0 GHz
50 MHz to 24 GHz
9 kHz to 6.0 GHz
60 dBm to
Maximum SWR
(25 °C±10 °C)
10 MHz to 30 MHz: 1.15
30 MHz to 2 GHz:
2 GHz to 14 GHz:
14 GHz to 16 GHz: 1.22
16 GHz to 18 GHz: 1.26
10 MHz to 30 MHz: 1.15
30 MHz to 2 GHz:
2 GHz to 6 GHz:
50 MHz to 2 GHz:
2 GHz to 14 GHz:
14 GHz to 16 GHz: 1.22
16 GHz to 18 GHz: 1.26
18 GHz to 24 GHz: 1.30
9 kHz to 2 GHz:
2 GHz to 6 GHz:
Introduction
The Agilent U2000 Series USB power sensors are average, dynamic wide-range
power sensors that can be used with a PC or any selected Agilent USB-based
instrument. With internal zeroing, there is no need to disconnect the sensor or
power off the device-under-test (DUT). The U2000 Series does not require any
50 MHz reference signal calibration, thus allowing factory calibration to ensure
measurement accuracy.
All the specifi cations provided in this data sheet are valid ONLY after proper
calibration of the power sensor and apply to continuous wave (CW) signals
unless otherwise stated.
The specifi cations are valid for a temperature range from 0 °C to +55 °C unless
otherwise stated. Specifi cations provided for a temperature of 25 °C ± 10 °C
are valid for a relative humidity range of 15% to 75% and conform to the
standard environmental test conditions.
The U2000 Series USB power sensors have two independent measurement
paths (high and low power paths) as shown below:
1.19
1.13
1.13
1.13
1.19
1.19
1.13
1.19
Maximum SWR (0-55 °C)
10 MHz to 30 MHz: 1.21
30 MHz to 2 GHz:
2 GHz to 14 GHz:
14 GHz to 16 GHz: 1.23
16 GHz to 18 GHz: 1.27
10 MHz to 30 MHz: 1.21
30 MHz to 2 GHz:
2 GHz to 6 GHz:
50 MHz to 2 GHz:
2 GHz to 14 GHz:
14 GHz to 16 GHz: 1.23
16 GHz to 18 GHz: 1.27
18 GHz to 24 GHz: 1.30
9 kHz to 2 GHz:
2 GHz to 6 GHz:
2
1.15
1.15
1.20
1.15
1.15
1.20
1.20
1.20
Power Range
–60 dBm to +20 dBm
Low Power Path: –60 dBm to –10 dBm
High Power Path: –10 dBm to +20 dBm
Maximum Power
+25 dBm (320 mW) average
+33 dBm peak (2 W) <10 us
+25 dBm (320 mW) average
+33 dBm peak (2 W) <10 us
+25 dBm (320 mW) average
+33 dBm peak (2 W) <10 us
+25 dBm (320 mW) average
+33 dBm peak (2 W) <10 us
Connector
Type
Type-N (m)
Type-N (m)
3.5 mm (m)
Type-N (m)