53230A Agilent Technologies Test Equipment, 53230A Datasheet

no-image

53230A

Manufacturer Part Number
53230A
Description
Universal Frequency Counter/Timer, 350 MHz, 12 digit/sec, 20 ps
Manufacturer
Agilent Technologies Test Equipment
Type
Counter/Timerr
Datasheet

Specifications of 53230A

Frequency Range
DC coupled 1 mHz to 350 MHz; AC coupled 10 Hz to 350 MHz
Input Impedance
Selectable 1 MΩ ±1.5% or 50Ω ±1.5%
Input Type
Front panel BNC
Input, Range
±5 V (± 50 V) full scale ranges
Noise
500 μVrms (max), 350 μVrms (typ)
Number Of Channels
3 Channels
Sensitivity
DC - 100 MHz: 20 mVpk > 100 MHz: 40 mVpk
53210A 350 MHz RF Frequency Counter, 10 digit/sec
53220A 350 MHz Universal Frequency Counter/Timer, 12 digit/sec, 100 ps
53230A 350 MHz Universal Frequency Counter/Timer, 12 digit/sec, 20 ps
Agilent 53200A Series
RF/Universal Frequency
Counter/Timers
Data Sheet

53230A Summary of contents

Page 1

... Agilent 53200A Series RF/Universal Frequency Counter/Timers Data Sheet 53210A 350 MHz RF Frequency Counter, 10 digit/sec 53220A 350 MHz Universal Frequency Counter/Timer, 12 digit/sec, 100 ps 53230A 350 MHz Universal Frequency Counter/Timer, 12 digit/sec ...

Page 2

... More Connectivity • LXI-C/Ethernet LAN, USB • Optional GPIB interface • Optional battery for portability and timebase accuracy More Measurement Capability (53230A only) • Continuous gap-free measurements • Basic modulation domain analysis (MDA) and timestamp • Optional pulse/burst microwave measurement Introduction Frequency counters are depended on in R& ...

Page 3

... Off: Selectable user set level (Volts) User selectable (Slow (50 Hz), Fast (10 kHz)) 300 mVpp 1 W Will not activate below 7.5 VpK 50 Ω internal termination auto-protects by switching Ω kHz: 350 Vpk (AC + DC) 5 kHz - 100 kHz: Derate linearly to 10 Vpk (AC + DC) >100 kHz: 10 Vpk (AC + DC) 3 53230A Ch 1 & ...

Page 4

... Autoranged to +19 dBm max. (2 Vrms) Autoranged to +15 dBm max. (1.25 Vrms) 6 GHz (Opt 106): -27 dBm (10 mVrms) 15 GHz (Opt 115): 0.3 GHz – 2 GHz: -23 dBm 2 GHz – 13 GHz: -26 dBm 13 GHz – 15 GHz: -21 dBm Auto-ranged for optimum sensitivity and bandwidth 50% modulation depth > +27 dBm (5 Vrms) 4 53230A ...

Page 5

... N/A Start delay (time or events) and stop hold-off N/A Independent start, stop slopes N/A Independent start, stop slopes N/A 100 ps 5 53230A optional digits/s 15 digits Reciprocal, resolution- enhanced or continuous (gap-free) CW and pulse/burst (Option 150) 1 µs to 1000 µs steps (time or events) ...

Page 6

... N/A Auto-level or user selectable N/A N/A N/A N/A N/A Continuous, timed, or external gate input ±5.1 Vpk with 2.5 mV resolution or ±51 Vpk with 25 mV resolution relative signal power 6 53230A - 100,000 s (nom) -0 100,000 s (min 100,000 s (min 1000 s Single-period: 2 -180.000º to 360.000º ...

Page 7

... N/A N Detector Level Burst carrier frequency + width burst (on width burst (off) PRF PRI = 1/PRF 7 53230A Carrier frequency, carrier period, pulse repetition interval (PRI), pulse repeti- tion frequency (PRF), posi- tive and negative width >200 ns Narrow: <17 us Wide: >13 us >50 ns – 1000 s ...

Page 8

... Rear panel BNC(f) Selectable as external gate input or gate output signal 50 Ω when selected for gate output TTL compatible Selectable positive or negative <-5 V, >+ 53230A Time, external or advanced 1 µs - 1000 s (1 µs) Internal or external, Ch 1/Ch 2 Positive or negative steps 8 for signals up to 100 MHz ...

Page 9

... Option 010 U-OCXO timebase) 1k Ω AC coupled 100 mVrms to 2.5 Vrms 10 MHz, 5 MHz, 1 MHz ±1 ppm (±0.1 ppm for Option 010 U-OCXO timebase) 7 Vrms 50 Ω ± MHz 0.5 Vrms into a 50 Ω load 1.0 Vrms into Ω load 10 MHz sinewave 7 Vrms 9 53230A 10 k/s max ...

Page 10

... Save & recall user-definable instrument setups Automatically saved Selectable power-on to reset (Factory), power-off state or user state 1 M readings (16 MBytes) 75 Mbytes ( readings) Front-panel connector for USB memory device Store/ recall user preferences and instrument states, reading memory, and bit map displays 10 53230A 2 , Max, Min, Peak-to-Peak, count ...

Page 11

... Maximum 53230A rates represent >= 20 MHz input signals with min gate times, no delays or holdoffs. Measurement rates for the 53210A & 53220A are limited by min gate time. Actual meas rates are limited by the repetition rate of the input being measured. ...

Page 12

... Complies with Australian standard and carries C-Tick Mark This ISM device complies with Canadian ICES-001 Cet appareil ISM est conforme a la norme NMB-001 du Canada SPL 35 dB(A) 100 V - 240 V 50/60 Hz -5%, +10% 100 V - 120 V 400 Hz ±10 max when powered On or charging battery max when powered off/standby 12 53230A ...

Page 13

... Warranty Standard 1. Assumes calibrated battery. 261.2 mm 103.8 mm 302.2 mm Dimensions apply to all three models: 53210A, 53220A, 53230A. continued 53210A 53220A Internal lithium ion battery with integrated smart battery monitor & charger 0 to 55º C Battery will only charge under 35º C Instrument running on battery power above 50º C will turn off to minimize battery capacity degradation. -10º ...

Page 14

Timebase Timebase Uncertainty = ( Aging + Temperature + Calibration Uncertainty ) Timebase 1 Aging (spec) 24-hour, T ±1º C CAL 30-day, T ±5º C CAL 1-year, T ±5º C CAL Temperature (typ) 0º 55º C relative to ...

Page 15

Accuracy Specifications Definitions Random Uncertainty The rss sum of all random or Type-A measurement errors expressed as the total RMS or 1-σ measurement uncertainty. Random uncertainty will reduce as 1/√N when averaging N measurement results for maximum ...

Page 16

... For 95% Confidence use k= 2.0 in accuracy calculations. continued Negative width, Negative duty, Single Channel Time Interval Fall to Rise Positive width, Positive duty, Single Channel Time Interval Rise to Fall 53210A 53220A 1 use R equation E 100 ps 100 ps 100 ps 200 and use gate as burst width. 53230A use R equation 100 ps ...

Page 17

Definition of Measurement Error Sources and Terms used in Calculations continued Resolution enhancement factor (R E The resolution enhancement (R ) calculates the added frequency resolution beyond the basic reciprocal measurement E capability that is achieved for a range of ...

Page 18

... Ordering Information Model numbers 53210A 350 MHz, 10-digit/s RF Frequency Counter 53220A 350 MHz, 12 digit/s, 100 ps Universal Frequency Counter/Timer 53230A 350 MHz, 12-digit/ Universal Frequency Counter/Timer All models include: • Shipping to your destination • Certificate of Calibration and 1-year standard warranty • IEC Power Cord, USB cable • ...

Page 19

Appendix A - Worked Example Basic Accuracy Calculation for Frequency Measurement Parameter assumptions: • 53220A • 95% Confidence • 100MHz signal, 1sec gate • AUTO • Level: 5V • OCXO standard timebase for unit plugged in for 30 days • ...

Page 20

... For more information on Agilent Technologies’ products, applications or services, please contact your local Agilent office. The complete list is available at: www.agilent.com/find/contactus Americas ...