U2001A Agilent Technologies Test Equipment, U2001A Datasheet

no-image

U2001A

Manufacturer Part Number
U2001A
Description
USB Power Sensor; 10 MHz - 6 GHz
Manufacturer
Agilent Technologies Test Equipment
Datasheet
Agilent
U2000 Series USB Power Sensors
Data Sheet

U2001A Summary of contents

Page 1

Agilent U2000 Series USB Power Sensors Data Sheet ...

Page 2

... Simplifi ed measurement setup with built-in triggering Table 1: Low power path and high power path for Agilent U2000 Series USB power sensors Sensor U2000A, U2001A, U2002A, U2004A Table 2: Product specifi cations for Agilent U2000 Series USB power sensors Model Frequency Range Maximum SWR (25 ° ...

Page 3

Why Zeroing Is Important For U2000 Series USB Power Sensor? Zeroing a power sensor is performed to reduce zero offset and noise impact to improve RF power measurement accuracy. The U2000 Series USB power sensor has two types of zeroing, ...

Page 4

... Typical SWR Chart (25 °C ±10 °C) Figure 2: Typical SWR chart for U2000A, 10 MHz to 18 GHz (25 °C ±10 °C) Figure 3: Typical SWR chart for U2001A, 10 MHz to 6 GHz (25 °C ±10 °C) 4 ...

Page 5

... Figure 4: Typical SWR chart for U2002A, 50 MHz to 24 GHz (25 °C ±10 °C) Figure 5: Typical SWR chart for U2001A, 9 kHz to 6 GHz (25 °C ±10 °C) 5 ...

Page 6

... Power Linearity Table 3: Power Linearity (after zero at ambient environment conditions) Sensor Power U2000A, U2001A, –60 dBm to +20 dBm U2002A, U2004A Power range –60 dBm to –35 dBm –38 dBm to –15 dBm –20 dBm to –9 dBm –11 dBm to –5 dBm –7 dBm to 15 dBm 10 dBm to 20 dBm Figure 6: Typical U2000 Series USB power sensors Power Linearity at 25 ° ...

Page 7

Switching Point The U2000 Series USB power sensors have two paths, a low power path covering –60 dBm to –10 dBm, and a high power path covering –10 dBm to +20 dBm. The power meter automatically selects the appropriate power ...

Page 8

Settling Time Settling Time for Normal Mode and x2 Mode Number of Averages 1 2 Settling Time1 (s) 0.045 0.09 (Normal Mode) Settling Time1 (s) 0.042 0.05 (x2 Mode) Noise Multiplier for Normal Mode and x2 Mode Number of Averages ...

Page 9

... GHz to 14 GHz ± 1.97% 14 GHz to 16 GHz ± 2.33% 16 GHz to 18 GHz ± 3.09% Table 6: Calibration factor uncertainties for U2001A Frequency –60 dBm to –10 dBm 10 MHz to 30 MHz ± 1.70% 30 MHz to 2 GHz ± 1.62% 2 GHz to 6 GHz ± ...

Page 10

Table 7: Calibration factor uncertainties for U2002A Frequency –60 dBm to –10 dBm ± 1.98% 50 MHz to 2 GHz 2 GHz to 14 GHz ± 2.27% 14 GHz to 16 GHz ± 2.34% 16 GHz to 18 GHz ± ...

Page 11

General Characteristics This instrument is designed for indoor use only. The table below shows the general requirements for the U2000 Series USB power sensors. Temperature 0 °C to +55 °C (operating) –30 °C to +70 °C (non-operating) Relative Humidity Operating ...

Page 12

Features Enhanced visualization with larger and more fl exible display formats Powerful graph functions Display multi-channel power measurements (with more than 10 displays on one window) Convenient data logging and storing up to seven days Time-saving options to save and ...

Page 13

Agilent Email Updates www.agilent.com/fi nd/emailupdates Get the latest information on the products and applications you select. Agilent Direct www.agilent.com/fi nd/agilentdirect Quickly choose and use your test equipment solutions with confi dence. Agilent Open www.agilent.com/fi nd/open Agilent Open simplifi es the ...