33220A Agilent Technologies Test Equipment, 33220A Datasheet

no-image

33220A

Manufacturer Part Number
33220A
Description
20 MHz Function/Arbitrary Waveform Generator, 14-Bit, 50 MSa/s, 10 Vpp
Manufacturer
Agilent Technologies Test Equipment
Datasheet

Specifications of 33220A

Frequency, Output
1 μHz to 20 MHz
Impedance, Output
50 Ohms (Typ.)
Temperature, Operating
0 to 55 °C
Voltage, Rating
10 mV to 10 VP-P
Uncompromising performance for functions
and waveforms
The Agilent Technologies 33220A Function/Arbitrary
Waveform Generator uses direct digital synthesis
(DDS) techniques to create a stable, accurate
output signal for clean, low distortion sine waves.
It also gives you square waves with fast rise and fall
times up to 20 MHz and linear ramp waves up to
200 kHz.
Pulse generation
The 33220A can generate variable-edge-time pulses
up to 5 MHz. With variable period, pulse width, and
amplitude the 33220A is ideally suited to a wide
variety of applications requiring a flexible pulse signal.
Custom waveform generation
Use the 33220A to generate complex custom
waveforms. With 14-bit resolution, and a sampling
rate of 50 MSa/s, the 33220A gives you the flexibility
to create the waveforms you need. It also lets you
store up to four waveforms in nonvolatile memory.
The Agilent IntuiLink Arbitrary Waveform software
allows you to easily create, edit, and download
complex waveforms using the waveform editor.
Or you can capture a waveform using IntuiLink
for Oscilloscope and send it to the 33220A for
output. To find out more about IntuiLink, visit
www.agilent.com/find/intuilink.
Easy-to-use functionality
Front-panel operation of the 33220A is straight-
forward and user friendly. You can access all major
functions with a single key or two. The knob or
numeric keypad can be used to adjust frequency,
amplitude, offset, and other parameters. You can
even enter voltage values directly in V
or as high and low levels. Timing parameters can be
entered in Hertz (Hz) or seconds.
Internal AM, FM, PM, FSK, and PWM modulation
make it easy to modulate waveforms without the
need for a separate modulation source. Linear and
logarithmic sweeps are also built in, with sweep
rates selectable from 1 ms to 500 s. Burst mode
operation allows for a user-selected number of
cycles per period of time. GPIB, LAN, and
USB interfaces are all standard, plus you get
full programmability using SCPI commands.
External frequency reference (Option 001)
The 33220A external frequency reference lets you
synchronize to an external 10 MHz clock, to another
33220A, or to an Agilent 33250A. Phase adjustments
can be made from the front panel or via a computer
interface, allowing precise phase calibration and
adjustment.
•20 MHz Sine and Square waveforms
•Pulse, Ramp, Triangle, Noise, and DC waveforms
•14-bit, 50 MSa/s, 64 K-point Arbitrary waveforms
•AM, FM, PM, FSK, and PWM modulation types
•Linear & logarithmic sweeps and burst operation
•10 mV
•Graph mode for visual verification of signal settings
•Connect via USB, GPIB and LAN
Agilent 33220A 20 MHz
Function/Arbitrary
Waveform Generator
Data Sheet
pp
to 10 V
pp
amplitude range
pp
, V
rms
, dBm,

33220A Summary of contents

Page 1

... Custom waveform generation Use the 33220A to generate complex custom waveforms. With 14-bit resolution, and a sampling rate of 50 MSa/s, the 33220A gives you the flexibility to create the waveforms you need. It also lets you store up to four waveforms in nonvolatile memory. The Agilent IntuiLink Arbitrary Waveform software allows you to easily create, edit, and download complex waveforms using the waveform editor ...

Page 2

WAVEFORMS Standard Sine, Square, Ramp, Triangle, Pulse, Noise, DC Built-in arbitrary Exponential rise, Exponential fall, Negative ramp, Sin(x)/x, Cardiac WAVEFORM CHARACTERISTICS Sine Frequency Range 1 µ MHz Amplitude Flatness [1], [2] (relative to 1 kHz) < 100 kHz ...

Page 3

... Trigger output TTL compatible into ≥ 1 kΩ Level Pulse width > 400 ns 50 Ω, typical Output Impedance Maximum rate 1 MHz ≤ 4 Agilent 33220As Fanout PROGRAMMING TIMES (typical) Configuration times USB Function Change 111 ms Frequency Change 1.5 ms Amplitude Change ...

Page 4

... Visit for more information. Agilent's IO Libraries Suite ships with the 33220A to help you quickly establish an error- free connection between your PC and instruments – regardless of the vendor. It provides robust instrument control and works with the software development environment you choose. ...