GPM140-30 SL Power, GPM140-30 Datasheet - Page 2

Linear & Switching Power Supplies 140W 100V-240V MED 30V 5.3A 1OUT

GPM140-30

Manufacturer Part Number
GPM140-30
Description
Linear & Switching Power Supplies 140W 100V-240V MED 30V 5.3A 1OUT
Manufacturer
SL Power
Series
GPMr
Datasheet

Specifications of GPM140-30

Product
Switching
Commercial/medical
Medical
Output Power Rating
140 W
Input Voltage
100 VAC to 240 VAC
Number Of Outputs
1
Output Voltage (channel 1)
30 V
Output Current (channel 1)
5.3 A
Mounting Style
Chassis
Open Frame/enclosed
Open Frame
Output Type
Single Output
circuits is 8 mm minimum. Secondary to ground creepage is not defined or controlled. The output common is bypassed to
ground using a 1 kV capacitor. The required creepage and clearance distances from primary circuits to ground and secondary
circuits must be maintained after installation to preserve the intended safety.
TEMPERATURES: The maximum operating temperatures of certain safety components, as defined in the applicable safety
standards, must not be exceeded after installation to preserve the intended safety. The output power, ambient air temperature
and the availability, amount, direction and/or restriction of airflow influence the temperatures of these components.
OVERCURRENT PROTECTION: The internal fuse is located in the phase lead only. EN 60601-1 requires that both
supply leads (phase and neutral) be protected against overcurrent. Complete overcurrent protection must be provided in the
host equipment. Fuse ratings must not exceed that specified for the internal fuse, must meet the requirements of EN 60601-1
and be acceptable for the country in which the host equipment is to be installed.
WARNING! RISK OF FIRE! A blown internal fuse is an indication of catastrophic failure of circuit component(s). Repair
must be performed by Condor authorized personnel. Refer to fuse marking on the supply for rating.
WARNING! SHOCK HAZARD! Dangerous voltages are present on some components, printed wiring traces and heatsinks.
Condor DC Power Supplies Inc. will not be liable for the safety, reliability or performance of these power supplies if a) any changes,
modifications or repairs are carried out by other than authorized agents of Condor DC Power Supplies Inc., or b) the installation of the
supply is not in accordance with these installation instructions and the applicable UL, CSA, and IEC/EN safety standards.
41-33340-0001 Rev. E 09/02/04
J1
1
3
5
TB1
MATING CONNECTORS:
1
2
3
CAUTION: Do not exceed 4 A per pin on J3 or 15 A per terminal on TB2.
Line
Neutral
Ground
AC INPUT
J2
1 Power Fail
2 - Sense
3 + Sense
4 N/C
EXPLANATION OF SYMBOLS
C O N N E C T I O N S
J1 Amp Housing 640250-5 Contact 770522-1
J2 Amp 640440-4
J3 Amp Housing 1-640250-6 Contact 770522-1
Alternating Current
Attention, Consult
Accompanying Documents
Attention, Dangerous Voltages
Protective Earth (Ground)
1,2,3,4,13,14,15,16
5,6,7,8,9,10,11,12
J3
1,2,7,8
3,4,5,6
TB2
DC OUTPUTS
+ Output
Return
Page 2 of 2

Related parts for GPM140-30