CUB4TC00 Red Lion Controls, CUB4TC00 Datasheet
![METER PANEL LCD THERMOCOUPLE](/photos/6/48/64884/mfg_cub4_series_sml.jpg)
CUB4TC00
Specifications of CUB4TC00
Q763911
CUB4TC00 Summary of contents
Page 1
... CUB4RT00 Overrange/Underrange Display: “…..”/”-…..” Normal Mode Rejection 50/60 Hz CUB4RT20 Common Mode Rejection: 100 CUB4TC00 Accuracies reflect ALL errors @ 12 VDC supply and 15 minute warm-up, except probe errors and lead resistance effects. CUB4TC20 MLPS1000 Note: Recommended minimum clearance (behind the panel) for mounting clip installation is 2.1" ...
Page 2
SPECIFICATIONS (Cont’d) CUB4TC Thermocouple Inputs: ACCURACY* ACCURACY* TC DISPLAY RANGE @ 23° 50°C TYPE ±°C -200 to 400° 2.3 -328 to 752°F -200 to 1372° 2.3 -328 to 2502°F -200 to 760°C J ...
Page 3
WIRING CONNECTIONS All conductors should meet voltage and current ratings of the unit. Also, cabling should conform to appropriate standards of good installation, local codes and regulations recommended that power supplied to the unit be protected by a ...
Page 4
This program item sets the display for Fahrenheit or Celsius and whole number or tenths. Tenths is not available for thermocouple types R, S, and B. To ...
Page 5
Voltage Calibration Diagram 1. Connect a precision Voltage Source with an output range of 0.000 to 105.000 mV and an accuracy of 0.03% or better to the SIG+ (#3) and SIG- (#5) terminals. Placement of jumper does not affect voltage ...
Page 6
VOLTAGE CALIBRATION Voltage Calibration Wiring 1. Connect a precision Voltage Source with an output range of -10.000 to 60.000 mV and an accuracy of 0.03% or better to the TC- (#5) and TC+ (#3) terminals. Cold junction calibration must be ...
Page 7
INSTALLATION The meter meets NEMA 4X/IP65 requirements for indoor use, when properly installed. The units are intended to be mounted on an enclosed panel. The viewing window and buttons are factory sealed for a wash down environment. A sponge rubber ...
Page 8
... The Company disclaims all liability for any affirmation, promise or representation with respect to the products. The customer agrees to hold Red Lion Controls harmless from, defend, and indemnify RLC against damages, claims, and expenses arising out of subsequent sales of RLC products or products containing components manufactured by RLC and based upon personal injuries, deaths, property ...