K3HB-XVA 24VAC/VDC Omron, K3HB-XVA 24VAC/VDC Datasheet - Page 34

no-image

K3HB-XVA 24VAC/VDC

Manufacturer Part Number
K3HB-XVA 24VAC/VDC
Description
PROCESS INDICATR AC VOLTAG INPUT
Manufacturer
Omron
Series
K3HBr
Type
Process Meterr
Datasheet

Specifications of K3HB-XVA 24VAC/VDC

Measuring Range
0 ~ 400V
Display Style
Green, Red Characters, Black Background
Display Type
LCD
Display Face Size
3.78" L x 1.89" W (96.0 x 48.0mm)
Display Digits
5
Display Digits - Height
0.193" (4.90mm), 0.559" (14.20mm)
Backlight
Green, Red
Mounting Type
Panel Mount
Termination
Screw Terminals
Voltage - Supply
24VAC/DC
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Other names
K3HB-XVA24VAC/VDC
K3HBXVA24VACVDC
21.Do not connect or remove connectors while the DeviceNet power
22.Use cables with a heat resistance of 70ºC min.
■ Noise Countermeasures
1. Do not install the product near devices generating strong high-fre-
2. Mount a surge suppressor or noise filter to peripheral devices
3. In order to prevent inductive noise, wire the lines connected to the
4. If a noise filter is used for the power supply, check the voltage and
5. Reception interference may occur if the product is used close to a
Example of Countermeasures for Inductive Noise on
Input Lines
is being supplied. Doing so will cause product failure or malfunc-
tion.
quency waves or surges, such as high-frequency welding and
sewing machines.
generating noise, in particular, motors, transformers, solenoids,
and magnet coils.
terminal block separately from power lines carrying high voltages
or currents. Do not wire in parallel with or in the same cable as
power lines. Other measures for reducing noise include running
lines along separate ducts and using shield lines.
current, and install the noise filter as close to the product as pos-
sible.
radio, television, or wireless.
Line filter
2 conductors
with shield
Indicator
Digital
Signal
input
-
+
Surge
suppressor
Indicator
Digital
Indicator
Digital
Panel Meter Series
K3HB
B-41