HI-8751 HANNA INSTRUMENTS, HI-8751 Datasheet - Page 4

THERMOMETER, PROBE TYPE

HI-8751

Manufacturer Part Number
HI-8751
Description
THERMOMETER, PROBE TYPE
Manufacturer
HANNA INSTRUMENTS
Datasheet

Specifications of HI-8751

Temperature Measuring Range
-40°C To +150°C
Temperature Tester Type
Thermometer - Electronic
External Height
180mm
External Width
83mm
External Depth
40mm
Accuracy
± 0.5% Full Scale
Range
Resolution
Accuracy
Typical EMC Deviation
Probe
Battery Type / Life
Environment
Dimensions
Weight
FUNCTIONAL DESCRIPTION & SPECIFICATIONS
1) Probe connector
2) Liquid Crystal Display
3) OFF mode
4) Measurement mode, °C scale
5) Measurement mode, °F scale
6) Rotary switch
1 x 9V (IEC 6LR61) / approx. 300 hours of continuous use
HI 765BL stainless steel, 1 m (3.3') cable (included)
±0.5% F.S. for one year, excluding probe error
180 x 83 x 40 mm (7.1x3.3x1.6")
–40.0 to 150.0°C / -58 to 338°F
OF HI 8753
RH max 95% non-condensing
0 to 50°C (32 to 122°F);
with HI 765BL probe
265 g (9.3 oz.)
±3°C / ±6°F
0.1°C / 1°F
6
Each meter is supplied complete with a 9V battery. Remove the battery cover
from the back of the thermometer, insert the battery and reattach the cover.
Connect the temperature probe to the probe socket on the top of the meter.
To switch the instrument on, turn the rotary switch to
°C or °F position (depending on model).
If a temperature probe is plugged in, the unit will display the measured
temperature.
After use, switch the thermometer off by turning the
rotary switch to OFF.
All HANNA instruments
the factory. It is generally recommended to have all thermometers recalibrated
at least once a year. For an accurate annual recalibration, contact the nearest
HANNA Customer Service Center.
For a quick and easy check of the meter accuracy,
HANNA instruments
TEST KEYS to be connected to the probe socket on
the thermometer.
If the reading differs more than ±0.4°C (±0.8°F) from the test key value, the
unit is due for recalibration.
Choose the test key that better fits your application needs:
HI 765-18C
HI 765000C
HI 765070C
HI 765-004F Calibration test key at
HI 765032F
HI 765158F
Calibration test key at –18.0°C
Calibration test key at
Calibration test key at
Calibration test key at
Calibration test key at
®
offers a series of CALIBRATION
OPERATIONAL GUIDE
®
thermometers have been accurately pre-calibrated at
CALIBRATION
7
158.0°F
– 0.4°F
70.0°C
32.0°F
0.0°C
OFF
OFF
°C
°C
°C
°C
°F
°F