OS534E NEWPORT ELECTRONICS, OS534E Datasheet - Page 31

INFRARED THERMOMETER, -18°C TO 871°C

OS534E

Manufacturer Part Number
OS534E
Description
INFRARED THERMOMETER, -18°C TO 871°C
Manufacturer
NEWPORT ELECTRONICS
Datasheet

Specifications of OS534E

Temperature Measuring Range
-18°C To 871°C
Temperature Tester Type
Thermocouple - Type K
Accuracy %
1%
Wavelength Max
670nm
2-16
LCK
2
2.3.7 Changing the Temperature from °F to °C (or vice
versa)
2.3.8 Turning on the Display Backlighting
2.3.9 Thermocouple Input (OS532E, OS533E, OS534E)
°F
Using the Handheld Infrared Thermometer
During the time the thermometer displays either d_F, d_M, MAX,
MIN, dIF, AVG or thermocouple temperature, press the
change all the temperatures from °F to °C or vice versa.
During the time that the thermometer displays either d_F, d_M,
LSR, MAX, MIN, dIF, AVG, or TC temperatures, press the
to turn the display backlighting ON/OFF.
The thermometer accepts thermocouple input. It displays
thermocouple temperature and the target temperature (via
infrared) simultaneously. This function provides an accurate
method of determining an unknown emissivity.
• To Determine an unknown target emissivity
1.
2.
3.
4.
5.
6.
Connect a contact thermocouple probe (Type K) to the
thermometer as shown in Figure 1-3.
Measure the object temperature using the thermocouple
probe.
Aim at the object and measure the temperature via infrared.
Press and hold the
mode (E) appears.
Set the emissivity by pressing the
the temperature reading matches the thermocouple
temperature measurement.
The thermometer now displays the correct object
emissivity.
"TC--- is” is displayed when the thermocouple
input is open or out of range (0 to 1600°F).
FUNC
key until the Emissivity Display
or
C
C
F
F
keys until
key to
key